Performances of four fourth‐generation human immunodeficiency virus‐1 screening assays
- 18 October 2012
- journal article
- research article
- Published by Wiley in Journal of Medical Virology
- Vol. 84 (12) , 1884-1888
Abstract
Fourth‐generation human immunodeficiency virus‐1 (HIV‐1) screening assays have improved sensitivity, but vary in performance characteristics. The purpose of this study was to evaluate four different fourth‐generation HIV‐1 assays. These assays included the AxSYM HIV Ag/Ab Combo (Abbott diagnostics, Delkenheim, Germany), ARCHITECT HIV Ag/Ab Combo (Abbott), Elecsys 2010 HIV Combi (Roche Diagnostics GmbH, Mannheim, Germany), and Elecsys HIV Combi PT (Roche). A total of 1,306 samples that included 1,225 clinical samples and 81 samples consisting of seroconversion panels, an HIV‐1 p24 antigen sensitivity panel, and dilution series of HIV‐1 lysates and HIV‐1 antibodies were tested. All of the assays had sensitivities of 100% on clinical samples. The specificities of the AxSYM, ARCHITECT, Elecsys 2010 HIV Combi, and Elecsys HIV Combi PT were 99.6, 99.6, 99.0, and 99.5%, respectively. Of the 81 samples with different levels of HIV antigen or antibody and/or subtypes, Elecsys HIV Combi PT and ARCHITECT HIV Ag/Ab Combo showed better analytical sensitivities than the other two assays. In summary, the performance characteristics of AxSYM, ARCHITECT, and Elecsys HIV Combi PT were comparable and satisfactory for clinical samples. ARCHITECT HIV Ag/Ab Combo and Elecsys HIV Combi PT have the higher analytical sensitivities, and would be preferable for reducing the window period. J. Med. Virol. 84:1884–1888, 2012.Keywords
This publication has 18 references indexed in Scilit:
- 2010 HIV Diagnostics ConferenceExpert Review of Anti-infective Therapy, 2010
- The risk of a second diagnostic window with 4th generation HIV assays: Two casesJournal of Clinical Virology, 2009
- Assessment of the Ability of a Fourth-Generation Immunoassay for Human Immunodeficiency Virus (HIV) Antibody and p24 Antigen To Detect both Acute and Recent HIV Infections in a High-Risk SettingJournal of Clinical Microbiology, 2009
- Performance of three automated fourth‐generation combined HIV antigen/antibody assays in large‐scale screening of blood donors and clinical samplesTransfusion Medicine, 2009
- Alternative Algorithms for Human Immunodeficiency Virus Infection Diagnosis Using Tests That Are Licensed in the United StatesJournal of Clinical Microbiology, 2008
- Could the new HIV combined p24 antigen and antibody assays replace p24 antigen specific assays?Journal of Virological Methods, 2007
- Performance evaluation of three automated human immunodeficiency virus antigen–antibody combination immunoassaysJournal of Virological Methods, 2006
- Comparative evaluation of the VIDAS HIV DUO Ultra assay for combined detection of HIV-1 antigen and antibodies to HIVJournal of Virological Methods, 2005
- Evaluation of the sensitivity and specificity of six HIV combined p24 antigen and antibody assaysJournal of Virological Methods, 2004
- Dynamics of HIV viremia and antibody seroconversion in plasma donorsAIDS, 2003