Quantitative analysis of treatment process time and throughput capacity for spot scanning proton therapy
- 9 June 2016
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 43 (7) , 3975-3986
Abstract
Purpose: To determine the patient throughput and the overall efficiency of the spot scanning system by analyzing treatment time, equipment availability, and maximum daily capacity for the current spot scanning port at Proton Therapy Center Houston and to assess the daily throughput capacity for a hypothetical spot scanning proton therapy center.Methods: At their proton therapy center, the authors have been recording in an electronic medical record system all treatment data, including disease site, number of fields, number of fractions, delivered dose, energy, range, number of spots, and number of layers for every treatment field. The authors analyzed delivery system downtimes that had been recorded for every equipment failure and associated incidents. These data were used to evaluate the patient census, patient distribution as a function of the number of fields and total target volume, and equipment clinical availability. The duration of each treatment session from patient walk‐in to patient walk‐out of the spot scanning treatment room was measured for 64 patients with head and neck, central nervous system, thoracic, and genitourinary cancers. The authors retrieved data for total target volume and the numbers of layers and spots for all fields from treatment plans for a total of 271 patients (including the above 64 patients). A sensitivity analysis of daily throughput capacity was performed by varying seven parameters in a throughput capacity model.Results: The mean monthly equipment clinical availability for the spot scanning port in April 2012–March 2015 was 98.5%. Approximately 1500 patients had received spot scanning proton therapy as of March 2015. The major disease sites treated in September 2012–August 2014 were the genitourinary system (34%), head and neck (30%), central nervous system (21%), and thorax (14%), with other sites accounting for the remaining 1%. Spot scanning beam delivery time increased with total target volume and accounted for approximately 30%–40% of total treatment time for the total target volumes exceeding 200 cm3, which was the case for more than 80% of the patients in this study. When total treatment time was modeled as a function of the number of fields and total target volume, the model overestimated total treatment time by 12% on average, with a standard deviation of 32%. A sensitivity analysis of throughput capacity for a hypothetical four‐room spot scanning proton therapy center identified several priority items for improvements in throughput capacity, including operation time, beam delivery time, and patient immobilization and setup time.Conclusions: The spot scanning port at our proton therapy center has operated at a high performance level and has been used to treat a large number of complex cases. Further improvements in efficiency may be feasible in the areas of facility operation, beam delivery, patient immobilization and setup, and optimization of treatment scheduling.Keywords
This publication has 14 references indexed in Scilit:
- Improving spot‐scanning proton therapy patient specific quality assurance with HPlusQA, a second‐check dose calculation engineMedical Physics, 2013
- Modelling the throughput capacity of a single-accelerator multitreatment room proton therapy centreBritish Journal of Radiology, 2012
- Impact of proton beam availability on patient treatment schedule in radiation oncologyJournal of Applied Clinical Medical Physics, 2012
- Is there a single spot size and grid for intensity modulated proton therapy? Simulation of head and neck, prostate and mesothelioma casesMedical Physics, 2012
- Quantitative analysis of beam delivery parameters and treatment process time for proton beam therapyMedical Physics, 2011
- Performance of the NIRS fast scanning system for heavy‐ion radiotherapyMedical Physics, 2010
- Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, HoustonMedical Physics, 2009
- The M. D. Anderson proton therapy systemMedical Physics, 2009
- Vision : Proton therapyMedical Physics, 2009
- Experiences at the Paul Scherrer Institute With a Remote Patient Positioning Procedure for High-Throughput Proton Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008