Abstract
Objective:In radiotherapy treatment, planning target volume and organs at risk are contoured on kilovoltage CT (kVCT) images. Unlike MR images, kVCT does not provide precise information on target volume extension. Since neither kVCT nor MRI may be suitable for contouring in patients with ferrous hip prostheses, this study evaluated whether megavoltage CT (MVCT) reduced interobserver variability.Methods:Two patients without hip prostheses and one patient (Patient 3) with hip prostheses were enrolled. Six radiation oncologists contoured prostate, rectum and bladder on kVCT (Patients 1 and 3), MRI (Patient 2) and MVCT images (Patient 3). MVCT was acquired with fine, normal and coarse modalities. Interobserver variability for each organ was analysed using conformity index (CI) and coefficient of variation (CV).Results:In patients without hip prostheses, CIs were higher in prostate contouring with MRI than with kVCT, indicating lower interobserver variability with MRI. Very slight variations were seen in rectu...

This publication has 19 references indexed in Scilit: