Abstract
Rationale: While the GC‐Orbitrap, marketed in 2015, represents a technological breakthrough in terms of sensitivity, resolution and mass stability, many studies have reported ion ratio modification in mass spectra using the standard 70 eV electron ionisation. Methods: We studied the influence of the acquisition and sample parameters leading to these modifications on fatty acid methyl esters (FAMEs). Results: FAMEs showed that these variations in relative intensities of ions were related to the acquisition parameters such as the mass range and the offset values of the C‐TRAP, but also directly related to the column concentration of the sample, and especially that it was molecule‐dependent. Advantageously, it is possible to use this feature to promote the molecular ions of FAMEs sometimes not present in a spectrum under electron ionisation at 70 eV. Conclusions: The 70 eV electron ionisation mass spectra from the GC‐Orbitrap were clearly molecule‐dependent and could be due to metastable ions during storage states in the C‐TRAP.