Fat Content Quantification with US Attenuation Coefficient: Phantom Correlation with MRI Proton Density Fat Fraction
Open Access
- 25 December 2025
- journal article
- research article
- Published by MDPI AG in Diagnostics
- Vol. 16 (1) , 80
Abstract
Objective: The aim of this study was to evaluate the consistency and reproducibility of attenuation coefficient (AC) measurements using different commercial ultrasound (US) across via a phantom experiment to investigate the relationship between the AC and MRI-derived proton density fat fraction (MRI-PDFF) values and the conversion equation. Methods: Twelve phantoms containing varying fat proportions (0–100%) were constructed. Phantom ACs were estimated via three US attenuation systems, including attenuation imaging (ATI), ultrasound attenuation analysis (USAT), and the US-guided attenuation parameter (UGAP), along with MRI-PDFF. Agreement among the AC values from the three ultrasonic attenuation systems was evaluated. Linear correlation analysis was used to explore the ACs, fat concentrations of the phantom, and MRI-PDFF measurements, from which a linear conversion formula between the ultrasonic attenuation parameters and the MRI-PDFF was derived. Results: MRI-PDFF and phantom fat concentration measurements appeared with a strong linear correlation (R2 = 0.996, p < 0.001). For the three US attenuation parameters, both inter-operator and intra-operator intraclass correlation coefficients (ICCs) ranged from 0.990 to 0.995 and 0.989 to 0.995, respectively. Bland–Altman analysis revealed no significant differences between the above three (all p > 0.05). Significant linear relationships were demonstrated between ultrasound attenuation parameters and phantom fat concentration (r = 0.938–0.986; all p < 0.001), as well as between ultrasound attenuation parameters and MRI-PDFF values (r = 0.922–0.982; all p < 0.001). A conversion formula (fat proportions ≤ 50%) was derived: US (dB/cm/MHz) = 0.501 + 0.012 MRI-PDFF (%). Conclusions: AC across different commercial ultrasound devices demonstrated significant diagnostic value in fat concentrations that appeared good consistency in measuring phantom fat concentration both between and within groups. The linear relationship between AC and MRI-PDFF enables the application of a conversion formula.Keywords
Funding Information
- National Natural Science Foundation of China (82302222)
- National Natural Science Foundation of China (82202191)
- Basic and Applied Basic Research Foundation of Guangdong Province (2023A1515220121)
- Natural Science Foundation of Guangdong Province (2024A1515011342)
This publication has 40 references indexed in Scilit:
- Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable ProtocolJournal of Visualized Experiments, 2018
- A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability ResearchJournal of Chiropractic Medicine, 2016
- Scatterer Number Density Considerations in Reference Phantom-Based Attenuation EstimationUltrasound in Medicine & Biology, 2014
- Multimodal Phantom of Liver TissuePLOS ONE, 2013
- Ultrasound Attenuation Measurements Using a Reference Phantom with Sound Speed MismatchUltrasonic Imaging, 2011
- Theoretical and phantom based investigation of the impact of sound speed and backscatter variations on attenuation slope estimationUltrasonics, 2011
- Validity of real time ultrasound in the diagnosis of hepatic steatosis: A prospective studyJournal of Hepatology, 2009
- Assessment of Hepatic Steatosis by Expert PathologistsAnnals of Surgery, 2009
- Design and validation of a histological scoring system for nonalcoholic fatty liver disease†Hepatology, 2005
- Tissue mimicking materials for ultrasound phantomsMedical Physics, 1978