INFLUENCE OF DIFFERENT TYPES OF PHANTOMS ON THE CALIBRATION OF DOSEMETERS FOR EYE LENS DOSIMETRY
Authors:
- 29 March 2016
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 170 (1-4) , 199-203
Abstract
Both a cylinder and a slab phantom have been recommended to be used as calibration phantoms for eye lens dosimetry in the International Atomic Energy Agency TECDOC. This study describes investigations on the influence of the type of phantom on the calibration of dosemeters. In order to fulfil the purpose, backscatter radiation from practically used water-filled phantoms was evaluated by calculations and experiments. For photons, the calculations showed that the cylinder phantom had 10 % lower backscattered effect at maximum than a slab phantom, and simulated well the backscattered effect of the human head or neck to within ±10 %. The irradiation results of non-filtered optically stimulated luminescence and radio-photoluminescence glass dosemeters indicated that the differences of the calibration factors between the two types of phantoms were up to 20 and 10 %, respectively, reflecting the response to backscattered photons. For electrons, no difference was found between the two types of phantoms.Keywords
This publication has 8 references indexed in Scilit:
- Particle and Heavy Ion Transport code System, PHITS, version 2.52Journal of Nuclear Science and Technology, 2013
- Backscattering factor for some personal dosimeters and impact on Hp(10)Radioprotection, 2013
- On the operational quantityHp(3) for eye lens dosimetryJournal of Radiological Protection, 2012
- Air kerma to Hp(3) conversion coefficients for a new cylinder phantom for photon reference radiation qualitiesRadiation Protection Dosimetry, 2012
- ICRP PUBLICATION 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection ContextAnnals of the ICRP, 2012
- Air kerma to dose equivalent conversion coefficients not included in ISO 4037-3Radiation Protection Dosimetry, 2010
- Introduction of the InLight monitoring serviceRadiation Protection Dosimetry, 2006
- Effect of Phantom Material on Backscattered Radiation against Photon Irradiation.RADIOISOTOPES, 2003