Bronchial rupture with a left‐sided polyvinylchloride double‐lumen tube
Authors:
- 15 April 2003
- journal article
- case report
- Published by Wiley in Acta Anaesthesiologica Scandinavica
- Vol. 47 (5) , 622-625
Abstract
Bronchial rupture after intubation with a double‐lumen endobronchial tube has been infrequently reported. Overinflation of the bronchial cuff was speculated to be a frequent cause of the bronchial damage. We report the case of a 78‐year‐old woman with non‐small cell carcinoma of the right upper lobe. Her trachea and left main‐stem bronchus were intubated with a left‐sided polyvinylchloride (PVC) double‐lumen endobronchial tube (Broncho‐Cath® 37 Fr, Mallinckrodth Medical, Athlone, Ireland). She underwent an uneventful right upper lobectomy. At the end of the resection, the surgeons noticed the herniating cuff from the ruptured left main‐stem bronchus. Laceration was repaired. Subsequent course of the patient was uneventful: she developed neither bronchial leak nor mediastinitis. Ten days later the patients was discharged home in a satisfactory condition. Factors that seem to increase the risk of injury by a double‐lumen tube are discussed.Keywords
This publication has 11 references indexed in Scilit:
- Airway rupture from double-lumen tubesJournal of Cardiothoracic and Vascular Anesthesia, 1999
- Placement of left-sided double-lumen endobronchial tubes: comparison of clinical and fibreoptic-guided placementBritish Journal of Anaesthesia, 1999
- Bronchial Rupture by a Double-Lumen Endobronchial Tube During Staging ThoracoscopyAnesthesia & Analgesia, 1999
- Predicting the Size of a Double-Lumen Endobronchial Tube Using Computed Tomographic Scan Measurements of the Left Main Bronchus DiameterAnesthesia & Analgesia, 1999
- A comparison of techniques for placement of double-lumen endobronchial tubesJournal of Clinical Anesthesia, 1998
- A Comparison of the Reliability of Two Techniques of Left Double-Lumen Tube Bronchial Cuff Inflation in Producing Water-Tight Seal of the Left Mainstem BronchusAnesthesia & Analgesia, 1998
- Predicting the Size of a Double-Lumen Endobronchial Tube Based on Tracheal DiameterAnesthesia & Analgesia, 1998
- Iatrogenic Ruptures of the Tracheobronchial TreeChest, 1997
- Fiberoptic bronchoscopy should not be a standard of care when positioning double-lumen endobronchial tubesJournal of Cardiothoracic and Vascular Anesthesia, 1994
- Bronchial Rupture Associated with the Use of a Double-Lumen Tube in a Small AdultAnesthesiology, 1989