Abstract
Summary: Focus stacking extracts the in-focus information from each image in a stack of focal planes and combines these in-focus selections into a single image representing an extended depth-of-field image with sharp, high-contrast anatomical details, where the entire field of view is in focus. In contrast, direct projections are more suited to focal sequences where out-of-focus information has already been removed by physical techniques such as X-ray μCT or confocal microscopy. While direct projections can also be applied to widefield focal stacks, the resulting image still contains out-of-focus information and is generally inferior to focus stacking. In this method review, software applications for extended depth-of-field using manual focus stacking are reviewed and compared to the direct projection of focal sequences. Several software algorithms are compared using examples of stained and unstained wood and bark based on manually acquired focal stacks. The advantages and disadvantages of different software applications are discussed in relation to ease of use and image quality. Focus stacking can be performed on manually acquired focus stacks with no special hardware and low-cost software, generating highly detailed images where the entire field of view is in focus, comparable to confocal microscopy but in widefield.