Dynamic T1-weighted magnetic resonance imaging of interstitial laser photocoagulation in the liver: Observations on in vivo temperature sensitivity

Authors:
  • PMPaul R. Morrison
    Paul R. Morrison
  • GKGilberto A. Kernahan
    Gilberto A. Kernahan

Abstract
Background and Objective Magnetic resonance imaging's (MRI) potential to monitor interstitial laser photocoagulation (ILP) has been previously demonstrated and is further tested here with improved spatial and temporal resolution. Study Design/Materials and Methods In vivo experiments employed fiber-delivered 1,064 nm light (3.0 W, 150 sec) in six rabbit livers monitored under T1-weighted FSE MRI as 1 image/10 sec and a 3 mm thick 8 cm FOV. Image signal intensities (SI) were compared with temperatures (T) at 7, 10, and 15 mm from the fiber. Results Data showed 33°C < T < 60°C. SI did not vary inverse-linearly with T; changes in the tissue altered the MRI signal interfering with the SI changes due to temperature. Conclusion MRI cannot map SI-derived temperatures over the entire treatment site. The role of MRI's temperature sensitivity must be coordinated with organ- and dose-specific tissue changes.