Microfluidics for the Rapid Detection of Pathogens Using Giant Magnetoresistance Sensors
- 2 December 2014
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Magnetics
- Vol. 50 (11) , 1-4
Abstract
This paper presents an integrated solution toward an on-chip microfluidic diagnostic system using the magnetically induced motion of functionalized magnetic microparticles (MPs) in combination with giant magnetoresistance (GMR) sensors. The innovative aspect of the proposed method is that the induced velocity on MPs in suspension, while imposed to a magnetic field gradient, is inversely proportional to their volume. Specifically, a velocity variation of suspended functionalized MPs inside a detection microchannel with respect to a reference velocity, specified in a parallel reference microchannel, indicates an increase in their nonmagnetic volume. This volumetric increase of the MPs is caused by the binding of pathogens (e.g., bacteria) to their functionalized surface. The new formed compounds, which have an increased nonmagnetic volume, are called loaded MPs (LMPs). Experiments with functionalized MPs and LMPs with Escherichia coli attached to their surface were conducted as a proof of concept. Their movement was demonstrated optically by means of a microscope with a mounted CCD camera as well as by measuring the resistance change of the integrated GMR sensors.Keywords
This publication has 13 references indexed in Scilit:
- On-chip microfluidic biosensor using superparamagnetic microparticlesBiomicrofluidics, 2013
- A Magnetic Biosensor System for Detection of E. coliIEEE Transactions on Magnetics, 2013
- A biodetection method using magnetic particles and micro trapsJournal of Applied Physics, 2012
- On-chip bio-analyte detection utilizing the velocity of magnetic microparticles in a fluidJournal of Applied Physics, 2011
- Advances in developing HIV-1 viral load assays for resource-limited settingsBiotechnology Advances, 2010
- Patient Needs and Point‐of‐Care Requirements for HIV Load Testing in Resource‐Limited SettingsThe Journal of Infectious Diseases, 2010
- Comparative evaluation of the VERSANT® HIV-1 RNA 1.0 kinetic PCR molecular system (kPCR) for the quantification of HIV-1 plasma viral loadJournal of Clinical Virology, 2009
- Cultivation gives context to the microbial ecologistFEMS Microbiology Ecology, 2007
- Magnetoresistive sensorsJournal of Physics: Condensed Matter, 2007
- Enzyme Immunoassay (EIA)/Enzyme-Linked Immunosorbent Assay (ELISA)Clinical Chemistry, 2005