Clinical evaluation of the CA530‐VET hematology analyzer for use in veterinary practice
Authors:
- Gisela ArndtGAGisela Arndt
- Bram BottemaBBBram Bottema
- Lothar JunkerLJLothar Junker
- Arthur GrabnerAGArthur Grabner
- Barbara KohnBKBarbara Kohn
- 1 June 2007
- journal article
- research article
- Published by Wiley in Veterinary Clinical Pathology
- Vol. 36 (2) , 155-166
Abstract
Background: The CA530‐VET is a completely automated impedance cell hematology analyzer, which yields a 16‐parameter blood count including a 3‐part leukocyte differential. Objectives: The aim of this study was to examine the operational potential of the CA530‐VET and its value for use in veterinary practice. Methods: The analyzer was tested for blood carry‐over, precision, and accuracy. Comparison methods included the CELL‐DYN 3500, microhematocrit centrifugation, manual platelet (PLT) counting for feline and equine species, and a 100‐cell manual WBC differential. Blood samples for comparison of the methods were obtained from 242 dogs, 166 cats, and 144 horses. Results: The carry‐over ratio (K) was 0.28% for RBC, 0.59% for PLT, 0.32% for WBC, and 0.18% for hemoglobin (HGB) concentration. Coefficients of variation (CVs) for within‐batch precision and duplicate measurement of blood samples were clearly within the required limits, except for duplicate platelet counts in cats (8.7%) and horses (9.5%). The WBC count was in excellent agreement for dogs and horses and RBC count was in excellent agreement for horses. The accuracy of feline WBC counts was not acceptable, with the exception of values at the high end of the range. RBC counts in dogs and cats, and HGB concentration and MCV in all 3 species were sufficiently accurate. The CA530‐VET HCT results were in excellent agreement with microhematocrit results in horses but exceeded the maximum allowed inaccuracy for cats and dogs. In all species, PLT counts established mechanically and manually were not in adequate agreement. Large differences were found between the CA530‐VET and the manual differential percentage for lymphocytes and “mid‐sized cells” (monocytes and basophilic granulocytes). Conclusions: The CA530‐VET can be considered useful for routine canine, feline, and equine blood cell analyses. It should not be considered accurate, however, for PLT counts, feline total WBC counts in the subnormal and normal range, and leukocyte differentials, except for granulocytes.Keywords
This publication has 10 references indexed in Scilit:
- Analysis of canine and feline haemograms using the VetScan HMT analyserJournal of Small Animal Practice, 2003
- Prevalence of Low Automated Platelet Counts in Cats: Comparison with Prevalence of Thrombocytopenia Based on Blood Smear EstimationVeterinary Clinical Pathology, 2001
- Evaluation of an in-house centrifugal hematology analyzer for use in veterinary practiceJournal of the American Veterinary Medical Association, 2000
- Evaluation of a Haematological Analyser (Sysmex F‐800) with Equine BloodJournal of Veterinary Medicine Series A, 1998
- Evaluation of a Hematology Analyzer With Canine and Feline BloodVeterinary Clinical Pathology, 1997
- Monocyte counting: discrepancies in results obtained with different automated instruments.Journal of Clinical Pathology, 1991
- Evaluation of a Haematological Analyser for its Use in Canine Clinical PathologyJournal of Veterinary Medicine Series A, 1991
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A Revised Scheme for the Evaluation of Automatic Instruments for Use in Clinical ChemistryAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1974
- Recommended scheme for the evaluation of instruments for automatic analysis in the clinical biochemistry laboratory.Journal of Clinical Pathology, 1969