Optimal flip angles for in vivo liver 3DT1 mapping and B1+ mapping at 3T
Open Access
- 1 May 2023
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 90 (3) , 950-962
Abstract
Purpose: The spoiled gradient recalled echo (SPGR) sequence with variable flip angles (FAs) enables whole liver mapping at high spatial resolutions but is strongly affected by inhomogeneities. The aim of this work was to study how the precision of acquired maps is affected by the and ranges observed in the liver at 3T, as well as how noise propagates from the acquired signals into the resulting map.Theory: The variance was estimated through the Fisher information matrix with a total noise variance including, for the first time, the map noise as well as contributions from the SPGR noise.Methods: Simulations were used to find the optimal FAs for both the mapping and mapping. The simulations results were validated in 10 volunteers.Results: Four optimized SPGR FAs of 2°, 2°, 15°, and 15° (TR = 4.1 ms) and map FAs of 65° and 130° achieved a coefficient of variation of 6.2 ± 1.7% across 10 volunteers and validated our theoretical model. Four optimal FAs outperformed five uniformly spaced FAs, saving the patient one breath‐hold. For the liver and parameter space at 3T, a higher return in precision was obtained by investing FAs in the SPGR acquisition rather than in the map.Conclusion: A novel framework was developed and validated to calculate the SPGR variance. This framework efficiently identifies optimal FA values and determines the total number of SPGR and measurements needed to achieve a desired precision.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/L016052/1)
- NIHR Oxford Biomedical Research Centre
This publication has 38 references indexed in Scilit:
- T 1 measurements in the human myocardium: The effects of magnetization transfer on the SASHA and MOLLI sequencesMagnetic Resonance in Medicine, 2013
- Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equationMagnetic Resonance in Medicine, 2008
- Rapid high‐resolution T1 mapping by variable flip angles: Accurate and precise measurements in the presence of radiofrequency field inhomogeneityMagnetic Resonance in Medicine, 2006
- T1, T2 relaxation and magnetization transfer in tissue at 3TMagnetic Resonance in Medicine, 2005
- Advances in functional and structural MR image analysis and implementation as FSLNeuroImage, 2004
- Determination of optimal angles for variable nutation proton magnetic spin‐lattice, T1, and spin‐spin, T2, relaxation times measurementMagnetic Resonance in Medicine, 2003
- Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady stateMagnetic Resonance in Medicine, 2003
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Imaging of the active B1 field in vivoMagnetic Resonance in Medicine, 1996
- Optimizing the precision in T1 relaxation estimation using limited flip anglesMagnetic Resonance in Medicine, 1987