Analyte Enrichment via Ion Concentration Polarization with Hydrogel Plugs Polymerized in PDMS Microchannels by a Facile and Comprehensive Method for Improved Polymerization
- 1 November 2022
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 94 (45) , 15586-15594
Abstract
Hydrogels are incorporated into microfluidic devices to provide enhanced functionality by enabling processes such as enzyme immobilization, sample enrichment, and ionic current rectification. However, in the microfluidic devices with the commonly used material poly(dimethylsiloxane) (PDMS), hydrogels are very difficult to polymerize in situ in an ambient atmosphere because of the high oxygen concentration in PDMS. Even with very high (1.8%) photoinitiator concentrations, the polymerized hydrogel does not completely fill the microchannel. Here, we report a facile and broadly applicable protocol that utilizes microchannel pretreatment with 20% benzophenone in acetone to provide a hydrogel plug that completely fills the microchannel cross section by consuming the oxygen in the PDMS substrate near the microchannel wall. Both negatively charged and neutral hydrogels are polymerized from monomer solutions that utilize the photoinitiator/solvent combinations of VA-086 in water and benzophenone or IRG in DMSO. The photoinitiators were tested at different concentrations and in devices with different levels of oxidation. The hydrogel morphology is characterized using phase contrast microscopy and is related to the hydrogel's performance for concentration enrichment and ionic current rectification. A novel method is employed to confine the precursor solution in desired locations so that a photomask is not required for the spatial control of the plug location. Among six hydrogel formulations, at 100 V, the best current rectification factor obtained is ∼600 and the best analyte enrichment achieved is ∼120-fold in 5 min. This method provides a rapid and simple approach to increase the capabilities of PDMS microfluidic devices through improved polymerization of nanoporous hydrogels.Keywords
Funding Information
- University of Pennsylvania
- National Institute of Biomedical Imaging and Bioengineering (R01EB025268)
This publication has 55 references indexed in Scilit:
- Spatiotemporal Mapping of Concentration Polarization Induced pH Changes at NanoconstrictionsACS Nano, 2012
- Benzophenone Absorption and Diffusion in Poly(dimethylsiloxane) and Its Role in Graft Photo-polymerization for Surface ModificationLangmuir, 2011
- Theory and experiments of concentration polarization and ion focusing at microchannel and nanochannel interfacesChemical Society Reviews, 2010
- Functionalized 3D-Hydrogel Plugs Covalently Patterned Inside Hydrophilic Poly(dimethylsiloxane) Microchannels for Flow-Through ImmunoassaysAnalytical Chemistry, 2009
- Ionic current rectification at a nanofluidic/microfluidic interface with an asymmetric microfluidic systemLab on a Chip, 2008
- Concentration Polarization and Nonlinear Electrokinetic Flow near a Nanofluidic ChannelPhysical Review Letters, 2007
- Electrokinetic concentration enrichment within a microfluidic device using a hydrogel microplugLab on a Chip, 2005
- Physics and Applications of Microfluidics in BiologyAnnual Review of Biomedical Engineering, 2002
- Role of oxygen in polymerization reactionsChemical Reviews, 1991