Physical image quality comparison of four types of digital detector for chest radiology
- 18 February 2008
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 129 (1-3) , 140-143
Abstract
Image quality for similar exposure conditions has been compared for two computed radiography (CR) systems (needle-based and conventional storage phosphor) and two flat-panel (DR) systems from different manufacturers mainly devoted to chest radiology. Image quality was assessed with a contrast-detail object and acrylic material to simulate clinical conditions. Specific image evaluation software was used to measure the contrast and obtain an image quality figure. Phantom and detector incident air kerma were measured for all images. Image quality differences were significant, and in the range of 100–300 µGy (phantom incident air kerma) the needle-based CR system and one of the DR systems show similar image quality and they are quite superior when compared with the conventional CR system.Keywords
This publication has 7 references indexed in Scilit:
- Transition from Screen-Film to Digital Radiography: Evolution of Patient Radiation Doses at Projection RadiographyRadiology, 2007
- Depiction of Low-Contrast Detail in Digital RadiographyInvestigative Radiology, 2006
- A Clinical Evaluation of the Image Quality Computer Program, CoCIQJournal of Digital Imaging, 2005
- Bildqualität und Dosis in der Digitalen ProjektionsradiographieRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2003
- Flat Panel Digital Radiography Compared With Storage Phosphor Computed RadiographyInvestigative Radiology, 2002
- Image quality vs radiation dose for a flat-panel amorphous silicon detector: a phantom studyEuropean Radiology, 2001
- Comparison of low-contrast detail perception on storage phosphor radiographs and digital flat panel detector imagesIEEE Transactions on Medical Imaging, 2001