Acute Evaluation of Programmer‐Guided AV/PV and VV Delay Optimization Comparing an IEGM Method and Echocardiogram for Cardiac Resynchronization Therapy in Heart Failure Patients and Dual‐Chamber ICD Implants
Authors:
- James H. BakerJBJames H. Baker
- John MckenzieJMJohn Mckenzie
- J. Russell BaileyJBJ. Russell Bailey
- 10 November 2006
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 18 (2) , 185-191
Abstract
Introduction: Intracardiac delay optimization of biventricular and dual‐chamber pacing devices currently relies on time‐consuming echocardiographic measurements. A novel intracardiac electrogram (IEGM) method for atrioventricular (AV/PV) and interventricular (VV) delay optimization was developed, which can be performed during routine device follow‐up. Methods and Results: In this prospective, nonrandomized, multi‐center trial, patients previously implanted with St. Jude Medical cardiac resynchronization therapy defibrillator (CRT‐D) devices or dual‐chamber implantable cardioverter defibrillators (ICDs) underwent standard AV/PV and/or VV delay optimization guided by Doppler echocardiogram measurements of the maximum aortic velocity time integral (aortic VTI). Aortic VTI measurements applying the IEGM method recommended delays were then obtained in all patients. Fifty‐eight patients (age: 68 ± 11 years; 81% male; 74% ischemic) and 57 patients (age: 71 ± 10 years; 74% male; 71% ischemic) were enrolled for AV/PV and VV delay evaluation, respectively. An independent core lab determined the maximum aortic VTIs.Data analysis of the AV, PV, and VV delays demonstrated the concordance correlation coefficient (CCC) between the standard method aortic VTI values and the IEGM method aortic VTI values was 97.5%, 96.1%, and 96.6%, respectively. All analyses demonstrated that the CCC > 90% (P < 0.05). Conclusion: The automated programmer‐based IEGM method provides a reliable and simpler alternative to standard techniques for the optimization of AV/PV and VV delay settings in patients with CRT‐D devices and dual‐chamber ICDs.Keywords
This publication has 24 references indexed in Scilit:
- Correlation of Echo-Doppler Optimization of Atrioventricular Delay in Cardiac Resynchronization Therapy With Invasive Hemodynamics in Patients With Heart Failure Secondary to Ischemic or Idiopathic Dilated CardiomyopathyThe American Journal of Cardiology, 2006
- Prospective comparison of echocardiographic atrioventricular delay optimization methods for cardiac resynchronization therapyHeart Rhythm, 2006
- Tailored echocardiographic interventricular delay programming further optimizes left ventricular performance after cardiac resynchronization therapyHeart Rhythm, 2005
- Understanding Nonresponders of Cardiac Resynchronization Therapy—Current and Future PerspectivesJournal of Cardiovascular Electrophysiology, 2005
- Echocardiographic Examination of Atrioventricular and Interventricular Delay Optimization in Cardiac Resynchronization TherapyThe American Journal of Cardiology, 2005
- Long‐Term Variations in Optimal Programming of Cardiac Resynchronization Therapy DevicesPacing and Clinical Electrophysiology, 2005
- Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapyThe American Journal of Cardiology, 2004
- Statistical Methods in Assessing AgreementJournal of the American Statistical Association, 2002
- Optimal atrioventricular delay at rest and during exercise in patients with dual chamber pacemakers: a non-invasive assessment by continuous wave Doppler.Heart, 1989
- Estimation of stroke volume changes by ultrasonic doppler.Circulation, 1977