Development of analytically capable time-of-flight mass spectrometer with continuous ion introduction
Authors:
- 1 March 2010
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 81 (3) , 033101
Abstract
The present article describes the results and findings explored in the course of the development of the analytically capable prototype of continuous time-of-flight (CTOF) mass spectrometer. Currently marketed pulsed TOF (PTOF) instruments use ion introduction with a 10 ns or so pulse width, followed by a waiting period roughly 100 μs. Accordingly, the sample is under excitation in 10−4 part of the total measuring time. This very low duty cycle severely limits the sensitivity of the PTOF method. A possible approach to deal with this problem is to use linear sinusoidal dual modulation technique (CTOF) as described in this article. This way the sensitivity of the method is increased, due to the 50% duty cycle of the excitation. All other types of TOF spectrometer use secondary electron multiplier (SEM) for detection, which unfortunately discriminates in amplification in favor of the lighter ions. This discrimination effect is especially undesirable in a mass spectrometric method, which targets high mass range. In CTOF method, SEM is replaced with Faraday cup detector, thus eliminating the mass discrimination effect. Omitting SEM is made possible by the high ion intensity and the very slow ion detection with some hundred hertz detection bandwidth. The electrometer electronics of the Faraday cup detector operates with amplification 1010 V/A. The primary ion beam is highly monoenergetic due to the construction of the ion gun, which made possible to omit any electrostatic mirror configuration for bunching the ions. The measurement is controlled by a personal computer and the intelligent signal generator Type Tabor WW 2571, which uses the direct digital synthesis technique for making arbitrary wave forms. The data are collected by a Labjack interface board, and the fast Fourier transformation is performed by the software. Noble gas mixture has been used to test the analytical capabilities of the prototype setup. Measurement presented proves the results of the mathematical calculations as well as the future potentiality for use in chemical analysis of gaseous mixtures.Keywords
This publication has 25 references indexed in Scilit:
- High-Resolution Electrospray Ionization Orthogonal-Injection Time-of-Flight Mass SpectrometerEuropean Journal of Mass Spectrometry, 2000
- Hadamard Transform Time-of-Flight Mass SpectrometryAnalytical Chemistry, 1998
- A Fourier transform time-of-flight mass spectrometer. A SIMION calculation approachInternational Journal of Mass Spectrometry and Ion Processes, 1995
- High-resolution mass spectrometry in a linear time-of-flight mass spectrometerInternational Journal of Mass Spectrometry and Ion Processes, 1994
- Laser assisted reflectron time-of-flight mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1994
- Reflectron time-of-flight mass spectrometry and laser excitation for the analysis of neutrals, ionized molecules and secondary fragmentsInternational Journal of Mass Spectrometry and Ion Processes, 1992
- Time-of-flight mass spectrometers with multiply reflected ion trajectoriesInternational Journal of Mass Spectrometry and Ion Processes, 1990
- The ideal mass analyzer: Fact or fiction?International Journal of Mass Spectrometry and Ion Processes, 1987
- A reflectron mass spectrometer with UV laser-induced surface ionizationInternational Journal of Mass Spectrometry and Ion Processes, 1987
- Fourier Transform Ion Mobility SpectrometryAnalytical Chemistry, 1985