Abstract
Background: Glucose oxidase (GOx)-based blood glucose monitors (BGMs) are influenced by the partial pressure of oxygen (P o 2 ) within the applied sample. Limited in-clinic data exists regarding the quantitative effect of P o 2 in unmanipulated capillary fingertip blood samples across physiologically representative glucose and P o 2 ranges. Method: Clinical accuracy data were collected as part of a BGM manufacturer’s ongoing post-market surveillance program for a commercially available GOx-based BGM test-strip. The data set comprised 29 901 paired BGM-comparator readings and corresponding P o 2 values from 5 428 blood samples from a panel of 975 subjects. Results: A linear regression-calculated bias range of 5.22% (+0.72% [low P o 2 : 45 mm Hg] to −4.5% [high P o 2 : 105 mm Hg]); biases calculated as absolute at 2 of 75 mm Hg, a linear regression bias of +3.14% was calculated at low P o 2 , while negligible impact on bias (regression slope: +0.002%) was observed at higher than nominal levels (>75 mm Hg). When evaluating BGM performance under corner conditions of low (180 mg/dL) glucose, combined with low and high P o 2 , linear regression biases ranged from +1.52% to −5.32% within this small group of subjects and with no readings recorded at 2 . Conclusions: Data from this large-scale clinical study, performed on unmanipulated fingertip capillary bloods from a diverse diabetes population, indicate P o 2 sensitivity of the BGM to be markedly lower than published studies, which are mainly laboratory-based, requiring artificial manipulation of oxygen levels in aliquots of venous blood.
Funding Information
  • LifeScan Scotland Ltd.