Clinical Accuracy of a Glucose Oxidase–Based Blood Glucose Test-Strip Across Extremes of Oxygen Partial Pressure
- 6 March 2023
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 18 (6) , 1445-1451
Abstract
Background: Glucose oxidase (GOx)-based blood glucose monitors (BGMs) are influenced by the partial pressure of oxygen (P o 2 ) within the applied sample. Limited in-clinic data exists regarding the quantitative effect of P o 2 in unmanipulated capillary fingertip blood samples across physiologically representative glucose and P o 2 ranges. Method: Clinical accuracy data were collected as part of a BGM manufacturer’s ongoing post-market surveillance program for a commercially available GOx-based BGM test-strip. The data set comprised 29 901 paired BGM-comparator readings and corresponding P o 2 values from 5 428 blood samples from a panel of 975 subjects. Results: A linear regression-calculated bias range of 5.22% (+0.72% [low P o 2 : 45 mm Hg] to −4.5% [high P o 2 : 105 mm Hg]); biases calculated as absolute at 2 of 75 mm Hg, a linear regression bias of +3.14% was calculated at low P o 2 , while negligible impact on bias (regression slope: +0.002%) was observed at higher than nominal levels (>75 mm Hg). When evaluating BGM performance under corner conditions of low (180 mg/dL) glucose, combined with low and high P o 2 , linear regression biases ranged from +1.52% to −5.32% within this small group of subjects and with no readings recorded at 2 . Conclusions: Data from this large-scale clinical study, performed on unmanipulated fingertip capillary bloods from a diverse diabetes population, indicate P o 2 sensitivity of the BGM to be markedly lower than published studies, which are mainly laboratory-based, requiring artificial manipulation of oxygen levels in aliquots of venous blood.Keywords
Funding Information
- LifeScan Scotland Ltd.
This publication has 5 references indexed in Scilit:
- Proof of Concept Study to Assess the Influence of Oxygen Partial Pressure in Capillary Blood on SMBG MeasurementsJournal of Diabetes Science and Technology, 2019
- How the reduction of O2 on enzymes and/or redox mediators affects the calibration curve of “wired” glucose oxidase and glucose dehydrogenase biosensorsElectrochimica Acta, 2013
- Partial Pressure of Oxygen in Capillary Blood Samples from the FingertipJournal of Diabetes Science and Technology, 2013
- Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testingCritical Care Medicine, 2001
- Ferrocene-mediated enzyme electrode for amperometric determination of glucoseAnalytical Chemistry, 1984