Verification of dosimetric accuracy on the TrueBeam STx: Rounded leaf effect of the high definition MLC
Authors:
- 1 October 2012
- journal article
- radiation therapy-physics
- Published by Wiley in Medical Physics
- Vol. 39 (10) , 6360-6371
Abstract
Purpose: The dosimetric leaf gap (DLG) in the Varian Eclipse treatment planning system is determined during commissioning and is used to model the effect of the rounded leaf‐end of the multileaf collimator (MLC). This parameter attempts to model the physical difference between the radiation and light field and account for inherent leakage between leaf tips. With the increased use of single fraction high dose treatments requiring larger monitor units comes an enhanced concern in the accuracy of leakage calculations, as it accounts for much of the patient dose. This study serves to verify the dosimetric accuracy of the algorithm used to model the rounded leaf effect for the TrueBeam STx, and describes a methodology for determining best‐practice parameter values, given the novel capabilities of the linear accelerator such as flattening filter free (FFF) treatments and a high definition MLC (HDMLC).Methods: During commissioning, the nominal MLC position was verified and the DLG parameter was determined using MLC‐defined field sizes and moving gap tests, as is common in clinical testing. Treatment plans were created, and the DLG was optimized to achieve less than 1% difference between measured and calculated dose. The DLG value found was tested on treatment plans for all energies (6 MV, 10 MV, 15 MV, 6 MV FFF, 10 MV FFF) and modalities (3D conventional, IMRT, conformal arc, VMAT) available on the TrueBeam STx.Results: The DLG parameter found during the initial MLC testing did not match the leaf gap modeling parameter that provided the most accurate dose delivery in clinical treatment plans. Using the physical leaf gap size as the DLG for the HDMLC can lead to 5% differences in measured and calculated doses.Conclusions: Separate optimization of the DLG parameter using end‐to‐end tests must be performed to ensure dosimetric accuracy in the modeling of the rounded leaf ends for the Eclipse treatment planning system. The difference in leaf gap modeling versus physical leaf gap dimensions is more pronounced in the more recent versions of Eclipse for both the HDMLC and the Millennium MLC. Once properly commissioned and tested using a methodology based on treatment plan verification, Eclipse is able to accurately model radiation dose delivered for SBRT treatments using the TrueBeam STx.Keywords
This publication has 16 references indexed in Scilit:
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- An Overview of Hypofractionation and Introduction to This Issue of Seminars in Radiation OncologySeminars in Radiation Oncology, 2008
- Dose deviations caused by positional inaccuracy of multileaf collimator in intensity modulated radiotherapyRadiation Measurements, 2008
- An experimental investigation into the radiation field offset of a dynamic multileaf collimatorPhysics in Medicine & Biology, 2006
- The use of film dosimetry of the penumbra region to improve the accuracy of intensity modulated radiotherapyMedical Physics, 2004
- Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committeeMedical Physics, 2003
- Calibration and quality assurance for rounded leaf‐end MLC systemsMedical Physics, 2001
- Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracyMedical Physics, 2000
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998