A novel method for patient exit and entrance dose prediction based on water equivalent path length measured with an amorphous silicon electronic portal imaging device
Authors:
- Awusi KavumaAKAwusi Kavuma
- Martin GleggMGMartin Glegg
- Mohamed MetwalyMMMohamed Metwaly
- Garry CurrieGCGarry Currie
- Alex ElliottAEAlex Elliott
- 17 December 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 55 (2) , 435-452
Abstract
In vivo dosimetry is one of the quality assurance tools used in radiotherapy to monitor the dose delivered to the patient. Electronic portal imaging device (EPID) images for a set of solid water phantoms of varying thicknesses were acquired and the data fitted onto a quadratic equation, which relates the reduction in photon beam intensity to the attenuation coefficient and material thickness at a reference condition. The quadratic model is used to convert the measured grey scale value into water equivalent path length (EPL) at each pixel for any material imaged by the detector. For any other non-reference conditions, scatter, field size and MU variation effects on the image were corrected by relative measurements using an ionization chamber and an EPID. The 2D EPL is linked to the percentage exit dose table, for different thicknesses and field sizes, thereby converting the plane pixel values at each point into a 2D dose map. The off-axis ratio is corrected using envelope and boundary profiles generated from the treatment planning system (TPS). The method requires field size, monitor unit and source-to-surface distance (SSD) as clinical input parameters to predict the exit dose, which is then used to determine the entrance dose. The measured pixel dose maps were compared with calculated doses from TPS for both entrance and exit depth of phantom. The gamma index at 3% dose difference (DD) and 3 mm distance to agreement (DTA) resulted in an average of 97% passing for the square fields of 5, 10, 15 and 20 cm. The exit dose EPID dose distributions predicted by the algorithm were in better agreement with TPS-calculated doses than phantom entrance dose distributions.Keywords
This publication has 28 references indexed in Scilit:
- Assessment of dosimetrical performance in 11 Varian a-Si500 electronic portal imaging devicesPhysics in Medicine & Biology, 2008
- Radiotherapy treatment verification using radiological thickness measured with an amorphous silicon electronic portal imaging device: Monte Carlo simulation and experimentPhysics in Medicine & Biology, 2008
- Comparison of three accelerated partial breast irradiation techniques: Treatment effectiveness based upon biological modelsRadiotherapy and Oncology, 2007
- Handbook of Radiotherapy Physics2007
- Investigation of an amorphous silicon EPID for measurement and quality assurance of enhanced dynamic wedgePhysics in Medicine & Biology, 2007
- Investigation of the optimal backscatter for an aSi electronic portal imaging devicePhysics in Medicine & Biology, 2004
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- The use of electronic portal imaging to verify patient position during intensity-modulated radiotherapy delivered by the dynamic MLC techniqueInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2002
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- A simple and robust method for in vivo midline dose map estimations using diodes and portal detectorsRadiotherapy and Oncology, 2001