Abstract
Purpose To evaluate endovascular repair of infrarenal abdominal aortic aneurysms (AAA) using the Endurant stent-graft and compare the outcomes of patients with different neck anatomies, particularly as pertains to stent-graft oversizing. Methods A retrospective review was conducted of 75 consecutive patients (69 men; mean age 75 years) undergoing endovascular AAA repair using the Endurant Stent Graft System from December 2008 to September 2011. The mean AAA size was 57± 10 mm (range 51–92), with a mean proximal neck length of 33±10 mm (9–127) and a mean infrarenal neck angulation of 25°±15° (0°–91°). Patients were stratified according to neck anatomy [reverse taper (n=22) vs. inside (n=44) and outside (n=9) the Instructions for Use (IFU) criteria]. Standard safety and efficacy outcome measures were augmented by measurements of the percent oversizing at the proximal and distal neck and volumes of the proximal neck and stent-graft. Results Technical success was 100% in all groups, with no early or late type Ia endoleak detected in any group. Procedure time, contrast volumes, and radiation dosages were comparable in all groups. The reverse taper neck group had stent-graft diameters and volumes that were significantly larger (p=0.007) than the other groups. The proximal neck oversizing of the endograft was significantly greater (p=0.008) in the reverse taper neck group (42.9%± 17.5%) compared to the within the IFU group (30.1%± 11.7%). Over a mean follow-up of 20 months (range 14–46), there were no aneurysm-related deaths and 9 type II endoleaks (5 in the reverse taper neck group; overall, 3 were treated and 6 resolved spontaneously). The outside the IFU group suffered no endoleak of any type and had no secondary interventions. Conclusion The Endurant stent-graft can be utilized with acceptable results in more challenging neck anatomies, such as those with a reverse taper, as long as there is adequate oversizing of the stent-graft.Keywords
This publication has 21 references indexed in Scilit:
- Thirty-day outcome and quality of life after endovascular abdominal aortic aneurysm repair in octogenarians based on the Endurant Stent Graft Natural Selection Global Postmarket Registry (ENGAGE)Journal of Vascular Surgery, 2012
- Early Results From a Canadian Multicenter Prospective Registry of the Endurant Stent Graft for Endovascular Treatment of Abdominal Aortic AneurysmsJournal of Endovascular Therapy, 2012
- Interdisciplinary and Translational Innovation: The Endurant Stent Graft…From Bedside to Benchtop and Back to BedsideJournal of Endovascular Therapy, 2011
- Final Results of the Prospective European Trial of the Endurant Stent Graft for Endovascular Abdominal Aortic Aneurysm RepairEuropean Journal of Vascular and Endovascular Surgery, 2011
- One-year multicenter results of 100 abdominal aortic aneurysm patients treated with the Endurant stent graftJournal of Vascular Surgery, 2011
- One year outcomes of the United States regulatory trial of the Endurant Stent Graft SystemJournal of Vascular Surgery, 2011
- Early results of the Endurant endograft system in patients with friendly and hostile infrarenal abdominal aortic aneurysm anatomyJournal of Vascular Surgery, 2011
- Evaluation of the Endurant stent graft under instructions for use vs off-label conditions for endovascular aortic aneurysm repairJournal of Vascular Surgery, 2011
- The Endurant Stent Graft System: 15-month follow-up report in patients with challenging abdominal aortic anatomiesLangenbeck's Archives of Surgery, 2011
- Endurant Stent-Graft: A 2-Year, Single-Center Experience With a New Commercially Available Device for the Treatment of Abdominal Aortic AneurysmsJournal of Endovascular Therapy, 2010