Inappropriate shock incidence in patients with subcutaneous implantable cardioverter‐defibrillators with concomitant cardiac implantable electronic devices: A single‐center cohort study
- 27 November 2023
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 47 (1) , 131-138
Abstract
BackgroundSubcutaneous implantable cardioverter defibrillators (S-ICDs) are occasionally used in combination with other cardiac implantable electronic devices (CIEDs). However, whether the incidence of inappropriate shock increases in patients with S-ICDs and concomitant CIEDs remains unclear. This study aimed to investigate the association between the concomitant use of CIEDs and the incidence of inappropriate shock in patients with current-generation S-ICDs.MethodsA total of 127 consecutive patients received an S-ICD. Patients were assigned to two groups depending on concomitant use of CIEDs at the time of S-ICD implantation: patients without other CIEDs (non-combined group, 106 patients) and patients with other CIEDs (combined group, 21 patients). CIEDs included pacemakers, implantable cardioverter defibrillators, cardiac resynchronization therapy pacemakers, and cardiac resynchronization therapy defibrillators. The primary outcome was inappropriate shock, defined as a shock other than ventricular arrhythmia. Hazard ratios and 95% confidence intervals were calculated using a time-varying Cox proportional hazards model which was adjusted for age because age differed between the groups and could be a confounder.ResultsDuring a median follow-up period of 2.2 years (interquartile range, 1.0-3.4 years), inappropriate shock events occurred in 17 (16%) and five (19%) patients of the non-combined and combined groups, respectively. While the age-adjusted hazard ratio for inappropriate shock was 24% higher in the combined than in the non-combined group (hazard ratio = 1.24, 95% confidence interval, 0.39-3.97), this difference was insignificant (p = .71).ConclusionThe incidence of inappropriate shock did not differ between patients with and without concomitant use of CIEDs, suggesting that S-ICDs could potentially be combined with other CIEDs without increasing the number of inappropriate shocks. Further studies are warranted to confirm the safety and feasibility of concomitant use of S-ICDs and CIEDs.Keywords
This publication has 20 references indexed in Scilit:
- Concomitant Use of the Subcutaneous Implantable Cardioverter Defibrillator and a Permanent PacemakerPacing and Clinical Electrophysiology, 2016
- Combining an subcutaneous ICD and a pacemaker with abdominal device location and bipolar epicardial left ventricular lead: first-in-man approachEP Europace, 2016
- Nothing inside the heart – Combining epicardial pacing with the S-ICDHeartRhythm Case Reports, 2015
- Combination of a leadless pacemaker and subcutaneous defibrillator: First in-human reportHeartRhythm Case Reports, 2015
- Subcutaneous implantable cardioverter-defibrillator: First single-center experience with other cardiac implantable electronic devicesHeart Rhythm, 2015
- A New Algorithm to Reduce Inappropriate Therapy in the S‐ICD SystemJournal of Cardiovascular Electrophysiology, 2015
- Effectiveness of Implantation of a Subcutaneous Implantable Cardioverter-Defibrillator in a Patient With Complete Heart Block and a PacemakerThe American Journal of Cardiology, 2015
- An Entirely Subcutaneous Implantable Cardioverter–DefibrillatorThe New England journal of medicine, 2010
- Immortal time bias in critical care research: Application of time-varying Cox regression for observational cohort studies*Critical Care Medicine, 2009
- Prognostic Importance of Defibrillator Shocks in Patients with Heart FailureNew England Journal of Medicine, 2008