Multilayer PDMS microfluidic chamber for controlling brain slice microenvironment
Open Access
- 24 May 2007
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 7 (7) , 842-849
Abstract
A novel three-layer microfluidic polydimethylsiloxane (PDMS) device was constructed with two fluid chambers that holds a brain slice in place with microposts while maintaining laminar perfusate flow above and below the slice. Our fabrication technique permits rapid production of PDMS layers that can be applied to brain slices of different shapes and sizes. In this study, the device was designed to fit the shape and thickness (530–700 µm) of a medullary brain slice taken from P0–P4 neonatal rats. Medullary slices in this chamber spontaneously produced rhythmic, respiratory-related motor output for up to 3 h, thereby demonstrating that brain slice viability was maintained for prolonged periods. This design is unique in that it achieves independent control of fluids through multiple channels in two separate fluid chambers. The laminar flow exhibited by the microfluidic chamber allows controlled solutions to target specific areas of the brain slice based on the input flow rates. To demonstrate this capability, a stream of Na+-free solution was focused on one half of a medullary slice to abolish spontaneous neural activity in only that half of the brain slice, while the other half remained active. We also demonstrated that flow of different solutions can be focused over the midline of the brain slice. The multilayer brain slice chamber design can integrate several traditional types of electrophysiology tools that are commonly used to measure neurophysiological properties of brain slices. Thus, this new microfluidic chamber is advantageous for experiments that involve controlled drug or solution delivery at high spatiotemporal resolution.Keywords
This publication has 27 references indexed in Scilit:
- Point:Counterpoint: The parafacial respiratory group (pFRG)/pre-Bötzinger complex (preBötC) is the primary site of respiratory rhythm generation in the mammalJournal of Applied Physiology, 2006
- Dynamic control of extracellular environment in in vitro neural recording systemsIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2005
- Continuous perfusion microfluidic cell culture array for high‐throughput cell‐based assaysBiotechnology and Bioengineering, 2004
- Silicon chip-based patch-clamp electrodes integrated with PDMS microfluidicsBiosensors and Bioelectronics, 2004
- Controlled microfluidic interfacesNature, 2004
- Microenvironment design considerations for cellular scale studiesLab on a Chip, 2004
- Physics and Applications of Microfluidics in BiologyAnnual Review of Biomedical Engineering, 2002
- Development of a microfluidic device for fluorescence activated cell sortingJournal of Micromechanics and Microengineering, 2002
- O2 tension in adult and neonatal brain slices under several experimental conditionsBrain Research, 1991
- The in vitro brain slice as a useful neurophysiological preparation for intracellular recordingJournal of Neuroscience Methods, 1980