Implementation and commissioning of an integrated micro-CT/RT system with computerized independent jaw collimation
Open Access
- 12 July 2013
- journal article
- radiation therapy-physics
- Published by Wiley in Medical Physics
- Vol. 40 (8) , 081706
Abstract
Purpose: To design, construct, and commission a set of computer‐controlled motorized jaws for a micro‐CT/RT system to perform conformal image‐guided small animal radiotherapy. Methods: The authors designed and evaluated a system of custom‐built motorized orthogonal jaws, which allows the delivery of off‐axis rectangular fields on a GE eXplore CT 120 preclinical imaging system. The jaws in the x direction are independently driven, while the y‐direction jaws are symmetric. All motors have backup encoders, verifying jaw positions. Mechanical performance of the jaws was characterized. Square beam profiles ranging from 2 × 2 to 60 × 60 mm2 were measured using EBT2 film in the center of a 70 × 70 × 22 mm3 solid water block. Similarly, absolute depth dose was measured in a solid water and EBT2 film stack 50 × 50 × 50 mm3. A calibrated Farmer ion chamber in a 70 × 70 × 20 mm3 solid water block was used to measure the output of three field sizes: 50 × 50, 40 × 40, and 30 × 30 mm2. Elliptical target plans were delivered to films to assess overall system performance. Respiratory‐gated treatment was implemented on the system and initially proved using a simple sinusoidal motion phantom. All films were scanned on a flatbed scanner (Epson 1000XL) and converted to dose using a fitted calibration curve. A Monte Carlo beam model of the micro‐CT with the jaws has been created using BEAMnrc for comparison with the measurements. An example image‐guided partial lung irradiation in a rat is demonstrated. Results: The averaged random error of positioning each jaw is less than 0.1 mm. Relative output factors measured with the ion chamber agree with Monte Carlo simulations within 2%. Beam profiles and absolute depth dose curves measured from the films agree with simulations within measurement uncertainty. Respiratory‐gated treatments applied to a phantom moving with a peak‐to‐peak amplitude of 5 mm showed improved beam penumbra (80%–20%) from 3.9 to 0.8mm. Conclusions: A set of computer‐controlled motorized jaws for a micro‐CT/RT system were constructed with position reliably better than a tenth of a millimeter. The hardware system is ready for image‐guided conformal radiotherapy for small animals with capability of respiratory‐gated delivery.Keywords
Funding Information
- Canadian Cancer Society (700386)
- Canadian Cancer Society (700386)
This publication has 26 references indexed in Scilit:
- A combined dose calculation and verification method for a small animal precision irradiator based on onboard imagingMedical Physics, 2012
- Small animal radiotherapy research platformsPhysics in Medicine & Biology, 2011
- Dosimetric characterization of an image-guided stereotactic small animal irradiatorPhysics in Medicine & Biology, 2011
- Characterization of image quality and image‐guidance performance of a preclinical microirradiatorMedical Physics, 2011
- Development of a Micro-Computed Tomography–Based Image-Guided Conformal Radiotherapy System for Small AnimalsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Image-guided small animal radiation research platform: calibration of treatment beam alignmentPhysics in Medicine & Biology, 2009
- High-Resolution, Small Animal Radiation Research Platform With X-Ray Tomographic Guidance CapabilitiesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- MicroRT—Small animal conformal irradiatorMedical Physics, 2007
- Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scannerMedical Physics, 2007
- Progress toward a microradiation therapy small animal conformal irradiatorMedical Physics, 2006