Baseline resolution of 1,2H and 3,4He isotope mixtures with a Paul ion trap mass spectrometer
Authors:
- 1 May 2003
- journal article
- letter
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 74 (5) , 2936-2938
Abstract
A Paul ion trap mass spectrometer with fractional resolution of was constructed and tested. The major improvement in this new trap over that previously reported with [Rev. Sci. Instrum. 73, 2157 (2002)] was due to tighter machining tolerances of the hyperboloidal surfaces, with minor improvement due to better stability of the amplitude of the radio frequency trapping voltage. The present resolution is shown to be sufficient to separate clearly (to baseline) the following masses: and (0.0256 u separation), and (0.0074 u), and (0.0059 u), and and (0.0143 u separation). Resolution of and (0.0043 u separation) should also be possible, but was not carried out. Resolution of and (0.0015 u separation) is possible with peaks of comparable intensity, or with deconvolution.
Keywords
This publication has 6 references indexed in Scilit:
- Real-time quantitative analysis of H2, He, O2, and Ar by quadrupole ion trap mass spectrometryJournal of the American Society for Mass Spectrometry, 2002
- A compact, high-resolution Paul ion trap mass spectrometer with electron-impact ionizationReview of Scientific Instruments, 2002
- Miniature, high-resolution, quadrupole mass-spectrometer arrayReview of Scientific Instruments, 1997
- Equilibrium space charge distribution in a quadrupole ion trapJournal of the American Society for Mass Spectrometry, 1994
- Ion trap mass spectrometer—past, present, and future (?)Mass Spectrometry Reviews, 1991
- Electromagnetic traps for charged and neutral particlesReviews of Modern Physics, 1990