Gradient system characterization by impulse response measurements with a dynamic field camera
Open Access
- 12 April 2012
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 69 (2) , 583-593
Abstract
This work demonstrates a fast, sensitive method of characterizing the dynamic performance of MR gradient systems. The accuracy of gradient time‐courses is often compromised by field imperfections of various causes, including eddy currents and mechanical oscillations. Characterizing these perturbations is instrumental for corrections by pre‐emphasis or post hoc signal processing. Herein, a gradient chain is treated as a linear time‐invariant system, whose impulse response function is determined by measuring field responses to known gradient inputs. Triangular inputs are used to probe the system and response measurements are performed with a dynamic field camera consisting of NMR probes. In experiments on a whole‐body MR system, it is shown that the proposed method yields impulse response functions of high temporal and spectral resolution. Besides basic properties such as bandwidth and delay, it also captures subtle features such as mechanically induced field oscillations. For validation, measured response functions were used to predict gradient field evolutions, which was achieved with an error below 0.2%. The field camera used records responses of various spatial orders simultaneously, rendering the method suitable also for studying cross‐responses and dynamic shim systems. It thus holds promise for a range of applications, including pre‐emphasis optimization, quality assurance, and image reconstruction. Magn Reson Med, 2013.Keywords
Funding Information
- Philips Healthcare
This publication has 29 references indexed in Scilit:
- Origin and reduction of motion and f0 artifacts in high resolution T2*-weighted magnetic resonance imaging: Application in Alzheimer's disease patientsNeuroImage, 2010
- Compensation of gradient-induced magnetic field perturbationsJournal of Magnetic Resonance, 2008
- Comments on ?Gradient-Induced Acoustic and Magnetic Field Fluctuations in a 4T Whole-Body MR Imager?Magnetic Resonance in Medicine, 2001
- An algorithm for eddy currents symmetrization and compensationMagnetic Resonance in Medicine, 1996
- A novel eddy current compensation scheme for pulsed gradient systemsMagnetic Resonance in Medicine, 1994
- Quantitative characterization of the eddy current fields in a 40‐cm bore superconducting magnetMagnetic Resonance in Medicine, 1994
- A simple method of measuring gradient induced eddy currents to set compensation networksMagnetic Resonance in Medicine, 1993
- Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: Optimization of corrections and quantitative characterization of magnet/gradient systemsMagnetic Resonance in Medicine, 1991
- Analytical method for the compensation of eddy-current effects induced by pulsed magnetic field gradients in NMR systemsJournal of Magnetic Resonance (1969), 1990