Determination of As, Mn, Pb and Se in coal by slurry electrothermal vaporization inductively coupled plasma mass spectrometry
Authors:
- 17 August 2000
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 15 (9) , 1081-1086
Abstract
A method for the determination of trace elements in coal by electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS), introducing the sample as a slurry, is proposed. The slurry was prepared by mixing the powdered coal (etc., were determined. By introducing 10 µL of a 4.0 mg mL−1 coal slurry, using 3 µg of Pd as a modifier, and a pyrolysis temperature of 600 °C, most of the obtained values for four coal reference materials were well within the certified range, using external calibration with aqueous solutions. No carrier had to be added in addition to the palladium modifier. Using a lower slurry concentration of 1.3 mg mL−1 and 2 µg of Pd as the modifier, the values obtained for lead were significantly lower than the certified ones. This problem could be solved in part by adding NaCl as a physical carrier, in addition to the palladium, showing that this carrier reduces the differences in the analyte transport from the sample and from the calibration solutions. However, results for most elements, except for arsenic, tended to be a little low even under these conditions, hence, preference was given to the higher slurry concentration.Keywords
This publication has 26 references indexed in Scilit:
- Influence of slurry preparation on the accuracy of ultrasonic slurry electrothermal atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1994
- Advances in ultrasonic slurry graphite furnace atomic absorption spectrometryAnalytical and Bioanalytical Chemistry, 1993
- SOLIDS ANALYSIS BY GFAASAnalytical Chemistry, 1992
- Solid sampling in electrothermal atomic absorption spectrometry using commercial atomizers. A reviewJournal of Analytical Atomic Spectrometry, 1991
- Direct atomic spectrometric analysis by slurry atomisation. Part 10. Use of an air-ashing stage in electrothermal atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1990
- Direct analysis of coal by electrothermal atomization atomic-absorption spectrometryTalanta, 1989
- Direct atomic spectrometric analysis by slurry atomisation. Part 7. Analysis of coal using inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1988
- Direct atomic spectrometric analysis by slurry atomisation. Part 4. Determination of selenium in coal by electrothermal atomisation atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1988
- Slurry sample preparation for simultaneous multi-element graphite furnace atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1988
- Applications of microwave oven sample dissolution in analysisAnalytical Chemistry, 1984