Optimized Reduced Field of View and Fat Suppression Methods for Interleaved Multislice In Vivo Cardiac Diffusion Tensor Imaging
Open Access
- 22 April 2026
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 96 (3) , 1097-1110
Abstract
Purpose: Slice interleaving, a limited phase encode (PE) field of view (FOV), and effective fat suppression are vital for efficient cardiac diffusion tensor imaging (cDTI) with minimal artifacts. This study aimed to optimize reduced FOV and fat suppression methods for interleaved multislice cDTI to improve signal‐to‐noise ratio (SNR) and minimize artifacts. Methods: Two‐slice motion compensated spin echo datasets from 20 healthy volunteers were acquired. Four reduced PE FOV sequences were evaluated: 2DRF pulse; applying either or pulses in PE direction; and the proposed flip‐back sequence with a nonselective pulse after readout to restore inverted magnetization. Four fat suppression techniques were implemented: no fat suppression (standard); fat saturation; binomial water excitation and spectral attenuated inversion recovery (SPAIR). Results: The proposed flip‐back sequence with SPAIR achieved the highest median SNR, and its SNR values are significantly higher () than 2DRF with SPAIR as current state‐of‐the‐art. SPAIR and water excitation demonstrated comparable performance when combined with the flip‐back sequence, and both yielded superior image quality than with no suppression or fat saturation. SPAIR showed robust fat suppression across most subjects, whilst water excitation exhibited advantages in some subjects with a high body mass index. Conclusion: The proposed flip‐back sequence with SPAIR enables efficient interleaved multislice imaging with reduced PE FOV and effective fat suppression, facilitating clinical translation of in vivo cDTI.Keywords
Funding Information
- British Heart Foundation (RG/F/23/110115)
- Siemens Healthineers (EP/S022104/1)
This publication has 53 references indexed in Scilit:
- Cardiac Diffusion: Technique and Practical ApplicationsJournal of Magnetic Resonance Imaging, 2019
- Diffusion tensor imaging of extraocular muscle using two-dimensional single-shot interleaved multiple inner volume imaging diffusion-weighted EPI at 3 teslaJournal of Magnetic Resonance Imaging, 2013
- Reduced field-of-view single-shot fast spin echo imaging using two-dimensional spatially selective radiofrequency pulsesJournal of Magnetic Resonance Imaging, 2010
- Fat and water magnetic resonance imagingJournal of Magnetic Resonance Imaging, 2009
- High‐resolution diffusion tensor imaging with inner field‐of‐view EPIJournal of Magnetic Resonance Imaging, 2009
- Reduction of eddy‐current‐induced distortion in diffusion MRI using a twice‐refocused spin echoMagnetic Resonance in Medicine, 2002
- Diffusion tensor imaging using single‐shot SENSE‐EPIMagnetic Resonance in Medicine, 2002
- Two‐dimensional spatially‐selective RF excitation pulses in echo‐planar imagingMagnetic Resonance in Medicine, 2002
- Improved clinical echo‐planar MRI using spatial‐spectral excitationJournal of Magnetic Resonance Imaging, 1998
- Adiabatic pulsesNMR in Biomedicine, 1997