Measurement of scintillator parameters using pulsed X-ray excitation
Authors:
- 1 August 2000
- journal article
- Published by Pleiades Publishing Ltd in Technical Physics Letters
- Vol. 26 (8) , 678-681
Abstract
A setup for measurement of the major scintillator parameters, including decay time constants, light yield, and emission spectrum, has been developed. It consists of three main components: a small-size source of short X-ray pulses, a cryostat, and a registration system. The ultimate parameters characterizing the source are as follows: pulse duration, ∼0.5 ns; repetition rate, 100 kHz; X-ray tube voltage, 30 kV; and current amplitude, 0.5 A. The registration system allows the measurements to be carried out in the 200–800 nm spectral range over the time intervals from 0.5 ns to 50 μs with resolution not worse than 0.1 ns.Keywords