Abstract
This article presents simulation and experimental results for add-on transmission attachments that have been developed for the scanning electron microscope(SEM). A primary beam energy of around 30 keV is used to irradiate thin test specimens (<100 nm). A transmission lens attachment enables a conventional field emissionSEM to improve its image resolution by an order of magnitude, providing an image magnification of several million and a spatial resolution down to 0.3 nm. In addition, a compact energy electron loss spectrometer attachment has been developed that uses the SEM’s built-in secondary electrondetector.