Steerable Guidewire for Magnetic Resonance Guided Endovascular Interventions
- 7 March 2014
- journal article
- Published by ASME International in Journal of Medical Devices
- Vol. 8 (2) , 021002
Abstract
In endovascular interventions, thin, flexible instruments are inserted through the skin into the blood vessels to diagnose and treat various diseases of the vascular system. One drawback is that the instruments are difficult to maneuver in the desired direction due to limitations in shape and flexibility. Another disadvantage is that the interventions are performed under intermittent fluoroscopy/angiography imaging. Magnetic resonance imaging (MRI) may offer advantages over X-ray guidance. It presents a good soft tissue contrast without the use of nephrotoxic media or ionizing radiation. The aim of this study is to develop a guidewire that is compatible with MRI and includes a steerable segment at the tip. This added degree-of-freedom may improve the maneuverability of the devices thereby the efficiently and safety of the navigation. A 1.6 m (5 ft, 3 in.) long and 0.035 in. diameter guidewire that consists of MR compatible materials and has a flexible tip was designed. The only metallic part was a nitinol rod that was implemented at the distal flexible tip. To limit the risk of heating in the MRI, this rod was kept shorter than 30 mm. The tip could be deflected in one direction by pulling on a Dyneema wire that was placed in the lumen of the shaft of the guidewire. To drive the steerable tip, a handle that could be easily attached/detached from the instrument was designed and implemented. Using the handle, the tip of the 1.60 m long guidewire prototype could be actuated to reach angles from 30 deg to 250 deg. The handle could easily be placed on and removed from the guidewire, so conventional 0.035 in.–compatible catheters could slide over from the proximal end. However, in order to make the guidewire more efficient to enter a bifurcation, the stiffness of the tip should progressively increase from its proximal to its distal end. The guidewire was imaged in a 1.5T MRI using real-time imaging without producing artifacts that would have shaded the anatomy. It was possible to assemble a guidewire with a steerable segment in the required size, using MR compatible materials. Therefore, the current design is a promising proof of concept and allowed us to clearly identify the features that need to be improved in order to come to a clinically applicable instrument.Keywords
This publication has 12 references indexed in Scilit:
- Steerable catheters in minimally invasive vascular surgeryThe International Journal of Medical Robotics and Computer Assisted Surgery, 2009
- Feasibility of Real-Time Magnetic Resonance–Guided Angioplasty and Stenting of Renal Arteries in Vitro and in Swine, Using a New Polyetheretherketone-Based Magnetic Resonance–Compatible GuidewireInvestigative Radiology, 2009
- MR-compatible polyetheretherketone-based guide wire assisting MR-guided stenting of iliac and supraaortic arteries in swine: Feasibility studyMinimally Invasive Therapy & Allied Technologies, 2009
- An MR guidewire based on micropultruded fiber‐reinforced materialMagnetic Resonance in Medicine, 2008
- Assessment of passive cardiovascular implant devices for MRI compatibilityTechnology and Health Care, 2008
- Surveillance of patients post-endovascular aneurysm repairPostgraduate Medical Journal, 2007
- Minimizing RF heating of conducting wires in MRIMagnetic Resonance in Medicine, 2007
- Safety of MRI‐guided vascular interventionsMinimally Invasive Therapy & Allied Technologies, 2006
- Cardiac catheterisation guided by MRI in children and adults with congenital heart diseaseThe Lancet, 2003
- Evaluation of Vascular and Interventional Procedures with Time–Action Analysis: A Pilot StudyJournal of Vascular and Interventional Radiology, 2002