Low-Current Measurements Using Transistor Logarithmic DC Electrometers
Author:
- 1 January 1970
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 17 (1) , 326-334
Abstract
A nine-decade (10-13 to 10-4 amp) and a five-decade (4 × 10-12 to 4 × 10-7 amp) dc logarithmic electrometer have been designed, and their performance evaluated. The problems of drift, linear-to-log conversion accuracy, temperature dependence, and frequency response are considered. An accuracy of better than ±0.6%/decade (~2% of reading) was achieved.Keywords
This publication has 4 references indexed in Scilit:
- A Circuit with Logarithmic Transfer Response over 9 DecadesIEEE Transactions on Circuit Theory, 1964
- Multiplication and Logarithmic Conversion by Operational Amplifier-Transistor CircuitsReview of Scientific Instruments, 1963
- AN Improved Solid-State Logarithmic AmplifierIEEE Transactions on Nuclear Science, 1963
- Semiconductor Logarithmic D. C. AmplifierIEEE Transactions on Nuclear Science, 1963