A simplified low‐cost phantom for image quality assessment of dental cone beam computed tomography unit
Open Access
- 15 November 2023
- journal article
- research article
- Published by Wiley in Journal of Medical Radiation Sciences
- Vol. 71 (1) , 78-84
Abstract
Introduction: A standardised testing protocol for evaluation of a wide range of dental cone beam computed tomography (CBCT) performance and image quality (IQ) parameters is still limited and commercially available testing tool is unaffordable by some centres. This study aims to assess the performance of a low-cost fabricated phantom for image quality assessment (IQA) of digital CBCT unit. Methods: A customised polymethyl methacrylate (PMMA) cylindrical phantom was developed for performance evaluation of Planmeca ProMax 3D Mid digital dental CBCT unit. The fabricated phantom consists of four different layers for testing specific IQ parameters such as CT number accuracy and uniformity, noise and CT number linearity. The phantom was scanned using common scanning protocols in clinical routine (90.0 kV, 8.0 mA and 13.6 s). In region-of-interest (ROI) analysis, the mean CT numbers (in Hounsfield unit, HU) and noise for water and air were determined and compared with the reference values (0 HU for water and -1000 HU for air). For linearity test, the correlation between the measured HU of different inserts with their density was studied. Results: The average CT number were -994.1 HU and -2.4 HU, for air and water, respectively and the differences were within the recommended acceptable limit. The linearity test showed a strong positive correlation (R-2 = 0.9693) between the measured HU and their densities. Conclusion: The fabricated IQ phantom serves as a simple and affordable testing tool for digital dental CBCT imaging.Keywords
This publication has 18 references indexed in Scilit:
- Quality assurance phantoms for cone beam computed tomography: a systematic literature reviewDentomaxillofacial Radiology, 2017
- Dental cone beam CT: A reviewPhysica Medica, 2015
- Quantitative performance characterization of image quality and radiation dose for a CS 9300 dental cone beam computed tomography machineJournal of Medical Imaging, 2015
- Effective dose of dental CBCT—a meta analysis of published data and additional data for nine CBCT unitsDentomaxillofacial Radiology, 2015
- Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantomDentomaxillofacial Radiology, 2013
- Development and implementation of a low-cost phantom for quality control in cone beam computed tomographyRadiation Protection Dosimetry, 2013
- Dental CBCT equipment and performance issuesRadiation Protection Dosimetry, 2012
- Development and applicability of a quality control phantom for dental cone‐beam CTJournal of Applied Clinical Medical Physics, 2011
- Optimization of exposure parameters for cone beam computed tomography sialographyDentomaxillofacial Radiology, 2011
- The effects of field-of-view and patient size on CT numbers from cone-beam computed tomographyPhysics in Medicine & Biology, 2009