Measuring bioelectric impedance outputs following coffee consumption in healthy adults using an 8-electrode segmental BIA device
Authors:
Open Access
- 3 July 2025
- journal article
- research article
- Published by Taylor & Francis in Journal of the International Society of Sports Nutrition
- Vol. 22 (1) , 2528531
Abstract
Bioelectric impedance analysis (BIA) is increasingly used to measure body composition in athletic, clinical and research settings. The reliability of BIA measurements relies on following procedures carefully. However, some practices for ensuring reliable measures may be unnecessarily restrictive. Previous research using BIA outputs as study outcome measures, has required participants to avoid coffee and caffeine-containing foods and beverages prior to measurements. The aim of this study was to determine whether BIA outputs are affected by coffee consumption at different caffeine concentrations. This study used a blinded, randomized, cross-over trial design. Participants (n = 13) received one of three treatments per visit: 200 mL hot water (80°C) mixed with 5 g instant coffee, 2.5 g instant coffee with 2.5 g decaffeinated coffee, or 5 g decaffeinated instant coffee. Body composition and fluid parameters were measured over 50 minutes using a Seca mBCA 515 device. The treatment predictor (p > 0.05) and sex-time-treatment interaction for all outcomes was found to be non-significant (p > 0.05). The time predictor was statistically significant (p < 0.05) for impedance, resistance and reactance but not for phase angle ϕ50 (p = 0.731), ϕ5 (p = 0.059) or urine osmolality (p = 0.066). The sex predictor was statistically significant for Z50 (p = 0.001), Z5 (p = 0.002), R50 (p = 0.001), R5 (p = 0.002), ϕ50 (p = 0.01), ϕ5 (p = 0.049), fat mass (%) (p = 0.016) and fat free mass (%) (p = 0.016). The effect size for this predictor was η2G < 0.336. A significant sex-time interaction was found for Z50 (p = 0.025) with a small effect size (η2G < 0.01). Effect sizes for the treatment predictor and time-treatment interaction were found to be small (η2G < 0.01). Effect size for the time predictor was small (η2G > 0.01). Changes in impedance, resistance and reactance were detected over the course of the experiment, and these changes were greater than could be explained by the technical error of measurement. However, the amount of caffeine in coffee did not affect BIA outputs. Effect sizes were small, suggesting little practical significance of drinking coffee before taking BIA measurements. Therefore, coaches, athletes, researchers and clinicians may be able to obtain reliable BIA measurements even when coffee has been consumed up to 50 minutes prior to measurements, however, fluid consumption and being adequately hydrated should still be considered.Keywords
Funding Information
- London Metropolitan University
This publication has 26 references indexed in Scilit:
- Impact of eating and drinking on body composition measurements by bioelectrical impedanceJournal of Human Nutrition and Dietetics, 2014
- Effect of Body Composition on Respiratory Compensation Point During an Incremental TestJournal of Strength & Conditioning Research, 2014
- No Evidence of Dehydration with Moderate Daily Coffee Intake: A Counterbalanced Cross-Over Study in a Free-Living PopulationPLOS ONE, 2014
- Assessing caffeine intake in the United Kingdom dietFood Chemistry, 2013
- Bioelectrical impedance analysis to estimate body composition in children and adolescents: a systematic review and evidence appraisal of validity, responsiveness, reliability and measurement errorObesity Reviews, 2013
- Notes on the History of Caffeine Use2010
- G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciencesBehavior Research Methods, 2007
- Bioelectrical impedance analysis in body composition measurement: National Institutes of Health Technology Assessment Conference StatementThe American Journal of Clinical Nutrition, 1996
- Body composition from fluid spaces and density: analysis of methods. 1961.1993
- Body Fat in Adult ManPhysiological Reviews, 1953