A study and optimization of lumbar spine X-ray imaging systems
Authors:
- M SandborgMSM Sandborg
- D R DanceDDD R Dance
- G Alm CarlssonGCG Alm Carlsson
- 1 March 2003
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 76 (903) , 177-188
Abstract
A Monte Carlo program has been developed that incorporates a voxel phantom of an adult patient in a model of the complete X-ray imaging system, including the anti-scatter grid and screen-film receptor. This allows the realistic estimation of patient dose and the corresponding image (optical density map) for a wide range of equipment configurations. This paper focuses on the application of the program to lumbar spine anteroposterior and lateral screen-film examinations. The program has been applied to study the variation of physical image quality measures and effective dose for changing system parameters such as tube voltage, grid design and screen-film system speed. These variations form the basis for optimization of these system parameters. In our approach to optimization, the best systems are those that can match (or come close to) the calculated image quality measure of systems preferred in a recent European clinical trial, but with lower patient dose. The largest dose savings found were 21% for a 400 speed class system with a grid having a strip density of 40 cm(-1) and a grid ratio of 16. A further dose saving of 13% was possible when a 600 speed class system was employed. The best systems found from the optimization correspond to those recommended by the European Commission guidelines on image quality criteria for diagnostic radiographic images.Keywords
This publication has 18 references indexed in Scilit:
- Schemes for the optimization of chest radiography using a computer model of the patient and x‐ray imaging systemMedical Physics, 2001
- The Optimisation of Lumbar Spine AP Radiography Using a Realistic Computer ModelRadiation Protection Dosimetry, 2000
- Comparison of Model Predictions of Image Quality with Results of Clinical Trials in Chest and Lumbar Spine Screen-film ImagingRadiation Protection Dosimetry, 2000
- A Voxel Phantom Based Monte Carlo Computer Program for Optimisation of Chest and Lumbar Spine X Ray Imaging SystemsRadiation Protection Dosimetry, 2000
- Image quality and dose in lumbar spine examinations: results of a 5 year quality control programme following the European quality criteria trialBritish Journal of Radiology, 1995
- Evaluation of image quality in fluoroscopy by measurements and Monte Carlo calculationsPhysics in Medicine & Biology, 1995
- Computerized three‐dimensional segmented human anatomyMedical Physics, 1994
- Voxel based Monte Carlo calculations of nuclear medicine images and applied variance reduction techniquesImage and Vision Computing, 1992
- Model of the spatial‐frequency‐dependent detective quantum efficiency of phosphor screensMedical Physics, 1990
- Random Noise Produced by X-Ray Fluorescent Screens*Journal of the Optical Society of America, 1968