Bench testing of a heterodyne CO2 laser dispersion interferometer for high temporal resolution plasma density measurements
- 1 December 2016
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 87 (12) , 123502
Abstract
A heterodyne detection scheme is combined with a 10.59 μm CO2 laser dispersion interferometer for the first time to allow large bandwidth measurements in the 10-100 MHz range. The approach employed utilizes a 40 MHz acousto-optic cell operating on the frequency doubled CO2 beam which is obtained using a high 2nd harmonic conversion efficiency orientation patterned gallium arsenide crystal. The measured standard deviation of the line integrated electron density equivalent phase resolution obtained with digital phase demodulation technique, is 4 × 1017 m−2. Air flow was found to significantly affect the baseline of the phase signal, which an optical table cover was able to reduce considerably. The heterodyne dispersion interferometer (DI) approach is found to be robustly insensitive to motion, with measured phase shifts below baseline drifts even in the presence of several centimeters of retroreflector induced path length variations. Plasma induced dispersion was simulated with a wedged ZnSe plate and the measured DI phase shifts are consistent with expectations.Keywords
Funding Information
- US-Japan Fusion Collaboration Program (FP5-3)
- U.S. Department of Energy (DE-FC02-06ER54875, DE-FC02-08ER54972)
This publication has 20 references indexed in Scilit:
- Development of dispersion interferometer for magnetic confinement plasmas and high-pressure plasmasJournal of Instrumentation, 2015
- Dispersion interferometer using modulation amplitudes on LHD (invited)Review of Scientific Instruments, 2014
- Application of phase-modulated dispersion interferometry to electron-density diagnostics of high-pressure plasmaJournal of Physics D: Applied Physics, 2014
- CO 2 laser imaging interferometer for high spatial resolution electron density profile measurements on LHDReview of Scientific Instruments, 2003
- Far infrared laser interferometer system on the Large Helical DeviceReview of Scientific Instruments, 1999
- Dual CO2 laser interferometer with a wavelength combination of 10.6 and 9.27 μm for electron density measurement on large tokamaksReview of Scientific Instruments, 1996
- Dispersion interferometer for controlled fusion devicesReview of Scientific Instruments, 1993
- Real-time, vibration-compensated CO2 interferometer operation on the DIII-D tokamakReview of Scientific Instruments, 1988
- MIRI: A multichannel far‐infrared interferometer/polarimeter for TFTR (abstract)Review of Scientific Instruments, 1985
- Second-harmonic interferometersOptics Letters, 1980