We developed a new source, having uniform spectral radiance, for a large-aperture UV integrating sphere with pressed polytetrafluoroethylene coating. Since lamp panel systems have high uncertainty due to a low signal-to-noise ratio at 250 nm, to transfer the spectral radiance scale from 250 nm to 350 nm to the source, we developed and calibrated a filter radiometer based on a solar blind phototube. The spectral radiance was obtained via a model of the source and a recursive estimation of the model parameters. A standard FEL lamp validated the usage of the filter radiometer. Finally, the filter radiometer to calibrate the new source. The results show that the radiance of the source at 250 nm is five times the top-of-the-atmosphere spectral irradiance, and the expanded uncertainty is 4.7%.Keywords
This publication has 15 references indexed in Scilit:
- In-lab calibration of space-borne differential optical absorption spectrometerOptics and Precision Engineering, 2013
- Radiometric comparison of the primary synchrotron radiation source standard Metrology Light Source with calibrated filter radiometers in the visible and NIR spectral rangeMetrologia, 2009
- Design and characterization of a large area uniform radiance source for calibration of a remote sensing imaging system2007
- The realization of the NIST detector-based spectral irradiance scaleMetrologia, 2003
- Spectral irradiance measurements of tungsten lamps with filter radiometers in the spectral range 290 nm to 900 nmMetrologia, 2000
- Nonlinearity measurements of silicon photodetectorsApplied Optics, 1998
- Development of a detector-based absolute spectral irradiance scale in the 380–900-nm spectral rangeApplied Optics, 1997
- An optical method for direct determination of the radiometric aperture area at high accuracyMeasurement Science and Technology, 1997
- Ultraviolet stability and contamination analysis of Spectralon diffuse reflectance materialOptical Engineering, 1993
- A new determination of the emissivity of tungsten ribbonPhysica, 1954