Detection of intraneural needle-placement with multiple frequency bioimpedance monitoring: a novel method
Authors:
Open Access
- 23 April 2015
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 30 (2) , 185-192
Abstract
Electrical impedance measurements have been used to detect intraneural needle placement, but there is still a lack of precision with this method. The purpose of the study was to develop a method for the discrimination of nerve tissue from other tissue types based on multiple frequency impedance measurements. Impedance measurements with 25 different frequencies between 1.26 and 398 kHz were obtained in eight pigs while placing the tip of a stimulation needle within the sciatic nerve and in other tissues. Various impedance variables and measurement frequencies were tested for tissue discrimination. Best tissue discrimination was obtained by using three different impedance parameters with optimal measurement frequencies: Modulus (126 kHz), Phase angle (40 kHz) and the Delta of the phase angle (between 126 and 158 kHz). These variables were combined in a Compound variable C. The area under the curve in a receiver operating characteristic was consecutively increased for the Modulus (78 %), Phase angle (86 %), Delta of the phase angle (94 %), and the Compound variable C (97 %), indicating highest specificity and sensitivity for C. An algorithm based on C was implemented in a real-time feasibility test and used in an additional test animal to demonstrate our new method. Discrimination between nerve tissue and other tissue types was improved by combining several impedance variables at multiple measurement frequencies.Keywords
This publication has 29 references indexed in Scilit:
- Ultrasound-Guided Evaluation of the Local Anesthetic Spread Parameters Required for a Rapid Surgical Popliteal Sciatic Nerve BlockRegional Anesthesia & Pain Medicine, 2010
- A finite element model of needle electrode spatial sensitivityPhysiological Measurement, 2010
- Invasive Electrical Impedance Tomography for Blood Vessel DetectionThe Open Biomedical Engineering Journal, 2010
- Impedance-based tissue discrimination for needle guidancePhysiological Measurement, 2009
- Pathophysiology of Peripheral Nerve Injury During Regional AnesthesiaRegional Anesthesia & Pain Medicine, 2008
- Pathophysiology of Peripheral Nerve Injury During Regional AnesthesiaRegional Anesthesia & Pain Medicine, 2008
- Quantitative Architecture of the Brachial Plexus and Surrounding Compartments, and Their Possible Significance for Plexus BlocksAnesthesiology, 2008
- Combination of intraneural injection and high injection pressure leads to fascicular injury and neurologic deficits in dogsRegional Anesthesia & Pain Medicine, 2004
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996
- The dielectric properties of biological tissues: I. Literature surveyPhysics in Medicine & Biology, 1996