A comparison of simulator-tested and -retrieved cervical disc prostheses
Authors:
- Paul A. AndersonPAPaul A. Anderson
- Jeffrey P. RouleauJRJeffrey P. Rouleau
- Jeffrey M. TothJTJeffrey M. Toth
- K. Daniel RiewKRK. Daniel Riew
- 1 September 2004
- journal article
- research article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery: Spine
- Vol. 1 (2) , 202-210
Abstract
Object. Total joint arthroplasties most commonly fail because the implant becomes worn and a host inflammatory response subsequently develops. Both the material response to the biological environment and the host response to the device must be thoroughly evaluated to establish the efficacy of cervical arthroplasty. Analyses of devices explanted in humans allow evaluation of both responses. Hypothetical wear rates can be determined by comparing in vivo wear with simulator-derived wear. The purpose of this study was to perform explant analyses involving the Bryan and Prestige discs and compare these results with those obtained using spine simulators. Methods. Of the approximately 5500 patients treated with the Bryan disc, 11 have undergone explantation of the device. Six of these devices were analyzed for dimensional and chemical changes. Three of the approximately 300 implanted Prestige discs were retrieved, and two were examined microscopically. Histological specimens were assessed for wear particles and host inflammatory response. Additionally, the extent of simulator-produced wear was compared with that demonstrated in the retrieved specimens. Conclusions. The simulator-generated results predict adequate wear-related characteristics for both the Bryan and Prestige prostheses for a minimum of 40 years. Comparison of data with those of the retrieved specimens indicates that the wear was more minimal than predicted in simulators by five- to 10-fold. In no instance did the revisions result from failure of the device due to a reaction to wear debris, fracture, polymer oxidation, or metal corrosion. The inflammatory response seen in the periprosthetic tissues was minimal and not characteristic of inflammatory responses in failed diarthrodial joint arthroplasties.Keywords
This publication has 15 references indexed in Scilit:
- Intermediate Follow-up After Treatment of Degenerative Disc Disease With the Bryan Cervical Disc Prosthesis: Single-Level and Bi-LevelSpine, 2003
- Wear Analysis of the Bryan Cervical Disc ProsthesisSpine, 2003
- The New Frenchay Artificial Cervical JointSpine, 2002
- Positive cytokine production in failed metal-on-metal total hip replacementsActa Orthopaedica, 2002
- SSE Up-DateEuropean Spine Journal, 2001
- Surgical experience with an implanted artificial cervical jointJournal of Neurosurgery, 1998
- Macrophage/particle interactions: Effect of size, composition and surface areaJournal of Biomedical Materials Research, 1994
- Histologic identification of polyethylene wear debris using Oil Red O stainJournal of Applied Biomaterials, 1993
- Age and Gender Related Normal Motion of the Cervical SpineSpine, 1992
- A biomechanical model for the analysis of the cervical spine in static posturesJournal of Biomechanics, 1991