Automatic noninvasive measurement of systolic blood pressure using photoplethysmography
Open Access
- 1 January 2009
- journal article
- research article
- Published by Springer Nature in BioMedical Engineering OnLine
- Vol. 8 (1) , 28
Abstract
Automatic measurement of arterial blood pressure is important, but the available commercial automatic blood pressure meters, mostly based on oscillometry, are of low accuracy. In this study, we present a cuff-based technique for automatic measurement of systolic blood pressure, based on photoplethysmographic signals measured simultaneously in fingers of both hands. After inflating the pressure cuff to a level above systolic blood pressure in a relatively slow rate, it is slowly deflated. The cuff pressure for which the photoplethysmographic signal reappeared during the deflation of the pressure-cuff was taken as the systolic blood pressure. The algorithm for the detection of the photoplethysmographic signal involves: (1) determination of the time-segments in which the photoplethysmographic signal distal to the cuff is expected to appear, utilizing the photoplethysmographic signal in the free hand, and (2) discrimination between random fluctuations and photoplethysmographic pattern. The detected pulses in the time-segments were identified as photoplethysmographic pulses if they met two criteria, based on the pulse waveform and on the correlation between the signal in each segment and the signal in the two neighboring segments. Comparison of the photoplethysmographic-based automatic technique to sphygmomanometry, the reference standard, shows that the standard deviation of their differences was 3.7 mmHg. For subjects with systolic blood pressure above 130 mmHg the standard deviation was even lower, 2.9 mmHg. These values are much lower than the 8 mmHg value imposed by AAMI standard for automatic blood pressure meters. The photoplethysmographic-based technique for automatic measurement of systolic blood pressure, and the algorithm which was presented in this study, seems to be accurate.Keywords
This publication has 13 references indexed in Scilit:
- Effects of External Pressure on Arteries Distal to the Cuff During SphygmomanometryIEEE Transactions on Biomedical Engineering, 2005
- Non-invasive measurement of systolic blood pressure on the arm utilising photoplethysmography: development of the methodologyMedical & Biological Engineering & Computing, 2005
- Extent of Cardiovascular Risk Reduction Associated With Treatment of Isolated Systolic HypertensionArchives of internal medicine (1960), 2003
- Isolated Systolic Hypertension: an Important Cardiovascular Risk FactorBlood Pressure, 1998
- Survey of Automated Noninvasive Blood Pressure MonitorsJournal of Clinical Engineering, 1994
- A practical and reliable method of measuring blood pressure in the neonate by pulse oximetryThe Journal of Pediatrics, 1994
- Update on Methods & Simulators for Evaluation of Noninvasive Blood Pressure MonitorsJournal of Clinical Engineering, 1994
- What blood pressure levels should be treated?Journal of Molecular Medicine, 1992
- Bilateral simultaneous indirect systolic blood pressure measurementsCardiovascular Research, 1986
- Assessment of Doppler ultrasound to measure systolic and diastolic blood pressures in infants and young children.Archives of Disease in Childhood, 1975