Although catheters with side holes allow high flow rate during hemodialysis, they also induce flow disturbances and create a critical hemodynamic environment that can favor fibrin deposition and thrombus formation. This study compared the blood flow and analyzed the influence of shear stress and shear rate in fibrin deposition and thrombus formation in nontunneled hemodialysis catheters with unobstructed side holes (unobstructed device) or with some side holes obstructed by blood thrombi (obstructed device). Computational fluid dynamics (CFD) was performed to simulate realistic blood flow under laminar and turbulent conditions. The results from the numerical simulations were compared with the fibrin distribution and thrombus architecture data obtained from scanning electron microscopy (SEM) and two photons laser scanning microscopy (TPLSM) on human thrombus formed in catheters removed from patients. CFD showed that regions of flow eddies and separation were mainly found in the venous holes region. TPLSM characterization of thrombi and fibrin structure in patient samples showed fibrin formations in accordance with simulated flux dynamics. Under laminar flow conditions, the wall shear stress close to border holes increased from 87.3 ± 0.2 Pa in the unobstructed device to 176.2 ± 0.5 Pa in the obstructed one. Under turbulent flow conditions, the shear stress increased by 47% when comparing the obstructed to the unobstructed catheter. The shear rates were generally higher than 5000/s and therefore sufficient to induce fibrin deposition. This findings were supported by SEM data documenting a preferential fibrin arrangement on side hole walls.Keywords
Funding Information
- Coordena��o de Aperfei�oamento de Pessoal de N�vel Superior (1014/11-0)
This publication has 21 references indexed in Scilit:
- Hemodialysis Catheter Thrombi: Visualization and Quantification of Microstructures and Cellular CompositionThe Journal of Vascular Access, 2013
- Architecture of fibrin network inside thrombotic material obtained from the right atrium and pulmonary arteries: flow and location matterJournal of Thrombosis and Thrombolysis, 2012
- Impact of altered venous hemodynamic conditions on the formation of platelet layers in thromboemboliThrombosis Research, 2012
- In vivo real-time two-photon microscopic imaging of platelet aggregation induced by selective laser irradiation to the endothelium created in the beta-actin-green fluorescent protein transgenic miceJournal of Thrombosis and Thrombolysis, 2011
- Effect of non-Newtonian and pulsatile blood flow on mass transport in the human aortaJournal of Biomechanics, 2011
- Composition of Coronary Thrombus in Acute Myocardial InfarctionJACC, 2011
- Internal-flow Nusselt numbers for the low-Reynolds-number end of the laminar-to-turbulent transition regimeInternational Journal of Heat and Mass Transfer, 2011
- Catheter Management in Hemodialysis PatientsClinical Journal of the American Society of Nephrology, 2011
- Spatial Distribution of Factor Xa, Thrombin, and Fibrin(ogen) on Thrombi at Venous ShearPLOS ONE, 2010
- Two-photon intravital imaging of thrombus developmentJournal of Biomedical Optics, 2010