Closed Loop Control of an MR-Conditional Robot with Wireless Tracking Coil Feedback
- 19 June 2019
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 47 (11) , 2322-2333
Abstract
This paper presents a hardware and software system to implement the task space control of an MR-conditional robot by integrating inductively coupled wireless coil based tracking feedback into the control loop. The main motivation of this work is to increase the accuracy performance and address the system uncertainties in the practical scenarios. We present the MR-conditional robot hardware design, wireless tracking method, and custom-designed communication software for real-time tracking data transfer. Based on these working principles, we fabricate the robot platform and evaluate the complete system with respect to various performance indices, i.e. data communication speed, targeting accuracy, tracking coil resolution, image quality, temperature variation, and task space control accuracy for static and dynamic targeting inside MRI scanner. The in-scanner targeting results show that the MR-conditional robot with wireless tracking coil feedback achieves the targeting error of 0.17 ± 0.08 mm, while the error calculated from the joint space optical encoder feedback is 0.68 ± 0.19 mm.Keywords
This publication has 30 references indexed in Scilit:
- Metal-related artifacts in instrumented spine. Techniques for reducing artifacts in CT and MRI: state of the artEuropean Spine Journal, 2009
- MRI‐compatible manipulator with remote‐center‐of‐motion controlJournal of Magnetic Resonance Imaging, 2008
- Pulse sequences and system interfaces for interventional and real‐time MRIJournal of Magnetic Resonance Imaging, 2008
- Actual flip‐angle imaging in the pulsed steady state: A method for rapid three‐dimensional mapping of the transmitted radiofrequency fieldMagnetic Resonance in Medicine, 2006
- MRI-Guided Biopsy of the Prostate Increases Diagnostic Performance in Men with Elevated or Increasing PSA Levels after Previous Negative TRUS BiopsiesEuropean Urology, 2006
- T1, T2 relaxation and magnetization transfer in tissue at 3TMagnetic Resonance in Medicine, 2005
- Integration of cardiac magnetic resonance imaging with three-dimensional electroanatomic mapping to guide left ventricular catheter manipulationJACC, 2004
- Applications of shape memory alloys in medical instrumentsMinimally Invasive Therapy & Allied Technologies, 2004
- Development of an MRI‐compatible catheter for pacing the heart: Initial in vitro and in vivo resultsJournal of Magnetic Resonance Imaging, 1996
- Rapid calculation of T1 using variable flip angle gradient refocused imagingMagnetic Resonance Imaging, 1987