Primary Apical Stability of Tapered Implants Through Reduction of Final Drilling Dimensions in Different Bone Density Models
- 1 December 2016
- journal article
- research article
- Published by Wolters Kluwer Health in Implant Dentistry
- Vol. 25 (6) , 775-782
Abstract
Purpose: A biomechanical study of the primary apical stability obtained in tapered implants through the reduction of final drilling dimensions in different bone density models. Material and Methods: An in vitro study of maximum insertion torque and primary stability based on the resonance frequency analysis (RFA) of 24 conical implants measuring 13 mm in length and 3.75 and 4.20 mm in diameter, randomly inserted in 10-mm sockets prepared in 4 polyurethane blocks with a density of 15, 20, 30, and 40 pounds per cu ft, respectively, reducing the diameter of the final drill at constant speed (400 rpm) to obtain exclusive 4 mm anchoring of the apical third of each implant. Results: The decrease in drilling diameter resulted in an increase in the insertion torque and implant stability quotient (ISQ) values in all implants, although without reaching statistical significance. In turn, a significant direct correlation was found between increasing bone analog block density and the insertion torque and ISQ values. Conclusions: Under the conditions of this study, the primary apical stability obtained may be more dependent on bone density than on reduction of the final drilling diameter.Keywords
This publication has 12 references indexed in Scilit:
- Biomechanical Evaluation of Undersized Drilling on Implant Biomechanical Stability at Early Implantation TimesJournal of Oral and Maxillofacial Surgery, 2013
- Effect of Drilling Dimension on Implant Placement Torque and Early Osseointegration Stages: An Experimental Study in DogsJournal of Oral and Maxillofacial Surgery, 2012
- CBCT fractal dimension changes at the apex of immediate implants placed using undersized drillingClinical Oral Implants Research, 2011
- Biomechanical Aspects of Primary Implant Stability: A Human Cadaver StudyClinical Implant Dentistry and Related Research, 2009
- Immediate Implant Placement: Clinical Decisions, Advantages, and DisadvantagesJournal of Prosthodontics, 2008
- Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implicationsPeriodontology 2000, 2008
- Stability and marginal bone level measurements of unsplinted implants used for mandibular overdentures: a 1‐year randomized prospective clinical study comparing early and conventional loading protocolsClinical Oral Implants Research, 2006
- Implant Stability during Initiation and Resolution of Experimental Periimplantitis: An Experimental Study in the DogClinical Implant Dentistry and Related Research, 2005
- Immediate Occlusal Loading the Same Day or the Day After Implant Placement: Comparison of 2 Different Time Frames in Totally Edentulous Lower JawsJournal of Oral Implantology, 2004
- Measurements Comparing the Initial Stability of Five Designs of Dental Implants: A Human Cadaver StudyClinical Implant Dentistry and Related Research, 2000