Suppress False Arrhythmia Alarms of ICU Monitors Using Heart Rate Estimation Based on Combined Arterial Blood Pressure and Ecg Analysis
Authors:
- Qiao LiQLQiao Li
- Gari D. CliffordGCGari D. Clifford
- 1 May 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21517614,p. 2185-2187
Abstract
Intensive care unit (ICU) monitors generate a high rate of false alarms when physiological signals are severely corrupted by noise. To suppress the false life-threatening heart rate (HR)-related electrocardiogram arrhythmia alarms, data derived from arterial blood pressure (ABP) signal were used. A new ABP signal quality index (SQI) was designed based upon the combination of two previously reported signal quality measures. HR was then tracked based on beat detection from ABP and a Kalman filter with a SQI-modified update sequence. Using the SQI to reject noisy ABP data, false HR-related arrhythmia alarms were rejected by comparing the estimated ABP-derived HR to the monitor's HR threshold. The algorithm was evaluated on 707 episodes of extreme bradycardia and 1877 episodes of extreme tachycardia alarms produced by a commercial ICU monitoring system. The false alarm reduction rate was 74.13% and 53.81%, and the corresponding true alarm acceptance rate was 99.60% and 99.58% for extreme bradycardia and extreme tachycardia respectively. Combining ECG and ABP information therefore provides a significant reduction in false alarms with minimal impact on true alarms.Keywords
This publication has 7 references indexed in Scilit:
- Robust heart rate estimation from multiple asynchronous noisy sources using signal quality indices and a Kalman filterPhysiological Measurement, 2007
- Reduction of false arterial blood pressure alarms using signal quality assessement and relationships between the electrocardiogram and arterial blood pressureMedical & Biological Engineering & Computing, 2004
- An open-source algorithm to detect onset of arterial blood pressure pulses2003
- MIMIC II: a massive temporal ICU patient database to support research in intelligent patient monitoring2002
- Poor prognosis for existing monitors in the intensive care unitCritical Care Medicine, 1997
- The median filter as a preprocessor for a patient monitor limit alarm system in intensive careComputer Methods and Programs in Biomedicine, 1991
- Physiologic trend detection and artifact rejection: a parallel implementation of a multi-state Kalman filtering algorithmComputer Methods and Programs in Biomedicine, 1990