Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement
- 19 January 2017
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 32 (1) , 13-22
Abstract
Intermittent non-invasive blood pressure measurement with tourniquets is slow, can cause nerve and skin damage, and interferes with other measurements. Invasive measurement cannot be safely used in all conditions. Modified arterial tonometry may be an alternative for fast and continuous measurement. Our aim was to compare arterial tonometry sensor (BPro®) with invasive blood pressure measurement to clarify whether it could be utilized in the postoperative setting. 28 patients who underwent elective surgery requiring arterial cannulation were analyzed. Patients were monitored post-operatively for 2 h with standard invasive monitoring and with a study device comprising an arterial tonometry sensor (BPro®) added with a three-dimensional accelerometer to investigate the potential impact of movement. Recordings were collected electronically. The results revealed inaccurate readings in method comparison between the devices based on recommendations by Association for the Advancement of Medical Instrumentation (AAMI). On a Bland–Altman plot, the bias and precision between these two methods was 19.8 ± 16.7 (Limits of agreement − 20.1 to 59.6) mmHg, Spearman correlation coefficient r = 0.61. For diastolic pressure, the difference was 4.8 ± 7.7 (LoA − 14.1 to 23.6) mmHg (r = 0.72), and for mean arterial pressure it was 11.18 ± 11.1 (LoA − 12.1 to 34.2) mmHg (r = 0.642). Our study revealed inaccurate agreement (AAMI) between the two methods when measuring systolic and mean blood pressures during post-operative care. The readings for diastolic pressures were inside the limits recommended by AAMI. Movement increased the failure rate significantly (p < 0.001). Thus, arterial tonometry is not an appropriate replacement for invasive blood pressure measurement in these patients.Keywords
Funding Information
- Pirkanmaan Rahasto (4/2014)
- Paulon Säätiö (12/2015)
- Suomen Anestesiologiyhdistys (2016)
This publication has 28 references indexed in Scilit:
- Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory useBlood Pressure Monitoring, 2013
- Arterial stiffness and pulse pressure in CKD and ESRDKidney International, 2012
- Comparison of two measuring instruments, B-pro and SphygmoCor system as reference, to evaluate central systolic blood pressure and radial augmentation indexHypertension Research, 2012
- Methods for the determination of limit of detection and limit of quantitation of the analytical methodsChronicles of Young Scientists, 2011
- A Comparison of Transmittance and Reflectance Pulse Oximetry During Vascular SurgeryAnesthesia & Analgesia, 2009
- Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac outputCritical Care, 2009
- The use of ambulatory tonometric radial arterial wave capture to measure ambulatory blood pressure: the validation of a novel wrist-bound device in adultsJournal of Human Hypertension, 2007
- An Accuracy Evaluation of the T-Line® Tensymeter (Continuous Noninvasive Blood Pressure Management Device) versus Conventional Invasive Radial Artery Monitoring in Surgical PatientsAnesthesia & Analgesia, 2006
- Recommendations for Blood Pressure Measurement in Humans and Experimental AnimalsHypertension, 2005
- On the Safety of Radial Artery CannulationAnesthesiology, 1983