Pump Versus Syringe: Aspiration Thrombectomy Direct Pressure Comparisons in a Comprehensive Benchtop 3-Dimensional-Printed Circle of Willis Model
- 10 June 2025
- journal article
- Published by Wolters Kluwer Health in Operative Neurosurgery
- Vol. 30 (2) , 224-232
Abstract
BACKGROUND AND OBJECTIVES: Experiments have shown that 60-cc syringes generate stronger vacuum forces than mechanical pumps. However, few studies have measured catheter tip suction pressure during thrombus engagement. Therefore, this study aimed to compare aspiration forces and pressure drops, generated by mechanical pumps and 60-cc syringes connected to various catheters when aspirating synthetic clots in a 3-dimensional-printed model. METHODS: The system includes a programmable pulsatile pump, a 3-dimensional-printed tissue-matched circle of Willis model, radiopaque synthetic blood clots positioned in the left internal carotid artery terminus, and pressure recording at 500 ± 10% mL/min and 120/80 ± 10 mm Hg. Each catheter underwent 3 soft and hard clot removal trials using a 60-cc syringe and an aspiration pump. Aspiration source pressure was recorded before, during, and after aspiration. The evaluated variables included engagement pressure change (∆P engagement ), aspiration pressure drops (∆P aspiration ), and pressure efficiency (P E ), defined as the percentage of peak pressure maintained during aspiration. RESULTS: Aspirating soft clots with syringes showed low engagement pressure drops (∆P engagement ) (0-14 mm Hg), high P E at 93%-99% of P p, and minimal aspiration pressure drops (∆P aspiration ) (6F catheters: 4-8 mm Hg, 8F catheters: 19 mm Hg). Conversely, using a pump generated moderate ∆P engagement (6F: 16-51 mm Hg, 8F: 122 mm Hg), lower P E (69%-93% of P p ), and greater ∆P aspiration (6F: 143-151 mm Hg, 8F: 270 mm Hg). For hard clots, syringes exhibited minimal ∆P engagement and ∆P aspiration (4-13 mm Hg and 3-12 mm Hg, respectively), whereas pumps showed moderate ∆P engagement (2-34 mm Hg) and large ∆P aspiration (6F: 82-162 mm Hg, 8F: 280 mm Hg). CONCLUSION: Syringe aspiration generates higher vacuum pressures and smaller pressure drops than aspiration pumps, likely because of vacuum tubing resistance in aspiration pumps. This effect is particularly evident with hard clots and larger inner diameter catheters.Keywords
This publication has 25 references indexed in Scilit:
- Recent development of aspiration pump in the aspiration thrombectomy systemMedicine in Novel Technology and Devices, 2024
- Clot composition and recanalization outcomes in mechanical thrombectomyJournal of NeuroInterventional Surgery, 2023
- Tensile and Compressive Mechanical Behaviour of Human Blood Clot AnaloguesAnnals of Biomedical Engineering, 2023
- Experimental evaluation of the performance of large bore aspiration cathetersJournal Of Neuroradiology, 2022
- Iatrogenic intracranial vessel dissection during mechanical thrombectomy rescued by emergent stenting: 2 case reportsRadiology Case Reports, 2021
- Clinical Comparison of New Generation 0.071-inch and 0.072-inch Aspiration CathetersWorld Neurosurgery, 2019
- Aspiration thrombectomy versus stent retriever thrombectomy as first-line approach for large vessel occlusion (COMPASS): a multicentre, randomised, open label, blinded outcome, non-inferiority trialThe Lancet, 2019
- A Direct Aspiration First Pass Technique vs Standard Endovascular Therapy for Acute Stroke: A Systematic Review and Meta-AnalysisNeurosurgery, 2017
- Endovascular thrombectomy after large-vessel ischaemic stroke: a meta-analysis of individual patient data from five randomised trialsThe Lancet, 2016
- Mechanical Thrombectomy for Acute Ischemic StrokeJACC, 2015