A survey of 29 EMI CT machines in Britain
Authors:
- R. D. SpellerRSR. D. Speller
- D. R. WhiteDWD. R. White
- R. VeerappanRVR. Veerappan
- A. J. ColemanACA. J. Coleman
- P.E. WallerPWP.E. Waller
- 1 March 1981
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 54 (639) , 232-240
Abstract
During 1978-1979 a survey was made of the performance characteristics of 29 EMI brain and body CT [computed tomography] scanners in Britain. Comparative data on such performance parameters as noise, uniformity of CT numbers, low-contrast detectability, resolution, modulation transfer function (MTF) and effective photon energy were measured using a set of regular-geometry and anthropomorphic phantoms. In addition, dose profiles were obtained for a single slice using TLD [thermoluminescent dosimetry] and film. The results of the survey are presented; the dependence of the performance characteristics on the types of machines surveyed and the operating modes employed is discussed.Keywords
This publication has 7 references indexed in Scilit:
- Evaluating performance characteristics in computerized tomographyBritish Journal of Radiology, 1981
- The measurement of effective photon energy and “linearity” in computerized tomographyBritish Journal of Radiology, 1980
- The radiation dose to patients from EMI brain and body scannersBritish Journal of Radiology, 1979
- Radiation exposure to the patient in computerized tomographyBritish Journal of Radiology, 1978
- Computerized tomography using the EMI general purpose scannerBritish Journal of Radiology, 1977
- The line spread function and modulation transfer function of a computed tomographic scannerMedical Physics, 1976
- Computerized transverse axial scanning (tomography): Part 1. Description of systemBritish Journal of Radiology, 1973