Predicting Transmural Lesion Formation and Steam‐Pop Occurrence During Bipolar Ablation—Ex Vivo Porcine Model
Open Access
- 7 April 2026
- journal article
- Published by Wiley in Journal of Arrhythmia
- Vol. 42 (2)
Abstract
Background: Bipolar radiofrequency (RF) ablation can create deeper myocardial lesions than unipolar ablation, yet its optimal settings remain undefined. Objective: To develop and validate predictive models for lesion transmurality and steam‐pop occurrence during bipolar ablation. Methods: Ex vivo bipolar ablation was applied to porcine myocardium (5–20 mm thickness) with catheters placed bilaterally at 45‐degree and 10‐g contact‐force. RF power (20–50 W) and duration (20–180 s) were systematically varied. Generalized linear models (GLM) were trained on 194 applications to predict transmurality and steam‐pop from RF energy, tissue thickness, initial bipolar impedance, 5‐s impedance drop (absolute and percentage [PercentImpDrop5]), and RF duration; 111 independent applications served as the validation. Results: Training yielded 95 transmural lesions (49%) and 11 steam‐pops (5.7%). For transmurality, the model incorporating RF energy, RF duration, initial impedance, and tissue thickness achieved an area under the receiver‐operating characteristics curve (AUC) of 0.95 (95% CI 0.91–0.99) with 88% sensitivity and 100% specificity. Omitting tissue thickness markedly degraded performance (AUC 0.68; DeLong, p = 0.003). For steam‐pop, the model combining RF energy and PercentImpDrop5 showed the best discrimination (AUC 0.90 [0.82–0.97], sensitivity 84%, specificity 90%); notably, PercentImpDrop5 alone achieved comparable accuracy (AUC 0.89). Conclusion: Tissue thickness is the dominant determinant of transmural lesion formation, whereas early impedance drop serves as a reliable real‐time indicator of steam‐pop risk during bipolar RF ablation. These algorithms may help standardize bipolar ablation protocols by enabling prospective titration of energy delivery based on tissue characteristics and intraprocedural impedance monitoring.Keywords
This publication has 14 references indexed in Scilit:
- Increasing Lesion Dimensions of Bipolar Ablation by Modulating the Surface Area of the Return ElectrodeJACC: Clinical Electrophysiology, 2022
- Bipolar ablation of therapy-refractory ventricular arrhythmias: application of a dedicated approachEP Europace, 2021
- Safety, efficacy, and monitoring of bipolar radiofrequency ablation in beating myopathic human and healthy swine heartsHeart Rhythm, 2021
- A case report: Amplatzer occluder device closure of an iatrogenic ventricular septal defect following radiofrequency ablationEuropean Heart Journal - Case Reports, 2021
- Impedance decrement indexes for avoiding steam‐pop during bipolar radiofrequency ablation: An experimental study using a dual‐bath preparationJournal of Cardiovascular Electrophysiology, 2020
- Higher contact force, energy setting, and impedance rise during radiofrequency ablation predicts charring: New insights from contact force‐guided in vivo ablationJournal of Cardiovascular Electrophysiology, 2017
- Bipolar irrigated radiofrequency ablation: A therapeutic option for refractory intramural atrial and ventricular tachycardia circuitsHeart Rhythm, 2012
- Steam pops during irrigated radiofrequency ablation: Feasibility of impedance monitoring for preventionHeart Rhythm, 2008
- The Biophysics of Radiofrequency Catheter Ablation in the Heart: The Importance of Temperature MonitoringPacing and Clinical Electrophysiology, 1993
- Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free WallPacing and Clinical Electrophysiology, 1989