Some causes for variation in aminoglycoside plate assays

Author:

Abstract
The reliability of gentamicin, tobramycin and netilmicin agar diffusion assays was closely associated with the slope of the standard curve, the influence of pH on the size of growth inhibition zone diameters and with the mode of applicating standard solutions and samples. The slope of the standard curve, expressed as the zone diameter difference obtained between twofold aminoglycoside dilutions, was determined by type and pH of the medium employed. Of 5 investigated media, Diagnostic Sensitivity Test agar (DST, Oxoid) yielded the most vavourable slope of standard curves and was also the least susceptible to discrepancies in pH between medium, standard solutions and samples. Wells filled to the brim with standard solutions and samples were associated with less intra-assay variation than wells and paper discs applied with set volumes. The DST agar well diffusion assays for gentamicin, tobramycin and netilmicin correlated well with a chemical assay involving high-pressure liquid chromatography and for gentamicin with a radioimmunoassay.

This publication has 1 reference indexed in Scilit: