Enumeration of feline platelets in ethylenediamine tetra-acetic acid anticoagulated blood with the ADVIA 2120 system and two manual methods
Open Access
- 18 June 2013
- journal article
- research article
- Published by SAGE Publications in Journal of Veterinary Diagnostic Investigation
- Vol. 25 (4) , 493-497
Abstract
A manual method (Thrombo-TIC; Bioanalytic GmbH, Umkirch/Freiburg, Germany) was advertised to disaggregate platelet clumps and to make human platelets spherical to improve platelet enumeration. The current study’s hypothesis was that this method would perform better than current methods for feline blood anticoagulated with ethylenediamine tetra-acetic acid (EDTA), which often contains platelet aggregates. Platelet concentrations (PLTs) were determined in 21 feline blood samples by 3 methods. Thrombo-TIC was compared to the manual method (Leucoplate; Sobioda, Montbonnot-Saint-Martin, France) currently used in the authors’ laboratory along with an ADVIA 2120 (Siemens AG, Eschborn, Germany) optical platelet concentration. Feline blood samples often contained platelet aggregates. ADVIA flagged for platelet aggregates in 11 of the 21 feline blood samples, and examination of blood smear revealed platelet aggregates in 14 of the 21 samples. The hemocytometers displayed more platelet aggregates with the Thrombo-TIC method than with Leucoplate. The method giving the greatest PLT was considered most accurate. The Leucoplate median PLT (238 × 109/1) was greater than Thrombo-TIC (202 × 109/1) or ADVIA (157 × 109/1). Intra-assay precision was determined for the 3 methods using the 21 feline blood samples. Median Thrombo-TIC and Leucoplate precision (7.4% and 7.3% coefficient of variation [CV], respectively) were similar and not much worse than ADVIA (5.9% CV). The Thrombo-TIC method did not appear to perform better than the current manual method (Leucoplate). Leucoplate appeared least affected by platelet aggregation in feline blood. The ADVIA automated PLT appeared to be most negatively affected by platelet aggregation. The Thrombo-TIC method did not appear to prevent platelet aggregation in feline EDTA blood samples and, thus, is not recommended for such use.Keywords
This publication has 8 references indexed in Scilit:
- Validation of Advia plateletcrit for assessing platelet mass in dogs, including Cavalier King Charles SpanielsVeterinary Clinical Pathology, 2012
- Feline platelet counting with prostaglandin E1 on the Sysmex XT‐2000iVVeterinary Clinical Pathology, 2010
- Validation of the Sysmex XT‐2000iV hematology system for dogs, cats, and horses. I. Erythrocytes, platelets, and total leukocyte countsVeterinary Clinical Pathology, 2009
- Performance Evaluation of the ADVIA 2120 Hematology Analyzer: An International Multicenter Clinical TrialLaboratory Hematology, 2005
- Prevalence of Low Automated Platelet Counts in Cats: Comparison with Prevalence of Thrombocytopenia Based on Blood Smear EstimationVeterinary Clinical Pathology, 2001
- Automated Analysis of Feline Platelets in Whole Blood, Including Platelet Count, Mean Platelet Volume, and Activation StateVeterinary Clinical Pathology, 1998
- Canine and feline haematology analysis: Comparative performance of technicon H*1 and AVL MS8 VET analysersComparative Clinical Pathology, 1994
- Thrombocytopenia in Cats: A Retrospective Study of 41 CasesJournal of Veterinary Internal Medicine, 1993