Evaluation of the dosimetric characteristics of 6 MV flattened and unflattened photon beam
Authors:
- Maged MohammedMMMaged Mohammed
- E. ChakirECE. Chakir
- H. BoukhalHBH. Boukhal
- S. MroanSMS. Mroan
- T. El BardouniTBT. El Bardouni
Open Access
- 1 July 2017
- journal article
- Published by Scientific Scholar in Journal of King Saud University – Science
- Vol. 29 (3) , 371-379
Abstract
This work aims to compare and evaluate the dosimetric properties of 6MV flattening (FF) and flattening filter free (FFF) photon beams which generated from a Varian 2100 medical accelerator. These properties include percentage depth dose, dose rate, beam profile, out-of-field, energy spectra, scatter factor, and surface dose. This study has been effected by using BEAMnrc and DOSXYZnrc user’s code based on EGSnrc Monte Carlo method.The results obtained showed that the unflattened beams have a dose rate of 2.46 times higher than the flattened beams that would reduce the treatment time. The out-of-field dose of FFF beams at 3.5cm from the field edge was less than the flattened one due to the reduction of head scatter. The scatter factor and penumbra dose of unflattened beam were found less than that of the flattened one for all field sizes. The unflattened beam has a higher surface dose and build-up dose compared to the flattened beams. Replacing the air column under the jaws by helium leads to the reduction of the surface dose ratio from 1.23 to 1.13, build-up doses and dmax restored as in flattened case. The results of this study have demonstrated that, unflattened beams are very useful for treating cancer cells and sparing the adjacent healthy tissueKeywords
This publication has 21 references indexed in Scilit:
- Comparison of VMAT‐SABR treatment plans with flattening filter (FF) and flattening filter‐free (FFF) beam for localized prostate cancerJournal of Applied Clinical Medical Physics, 2015
- The effect of flattening filter free delivery on endothelial dose enhancement with gold nanoparticlesMedical Physics, 2013
- Commissioning of Photon Beams of a Flattening Filter-Free Linear Accelerator and the Accuracy of Beam Modeling Using an Anisotropic Analytical AlgorithmInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Current status and future perspective of flattening filter free photon beamsMedical Physics, 2011
- Dosimetric characteristics of 6 and 10MV unflattened photon beamsRadiotherapy and Oncology, 2009
- Frequency Distribution of Second Solid Cancer Locations in Relation to the Irradiated Volume Among 115 Patients Treated for Childhood CancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Dosimetric characteristics of unflattened 6MV photon beams of a clinical linear accelerator: A Monte Carlo studyApplied Radiation and Isotopes, 2007
- Reduced Neutron Production Through Use of a Flattening-Filter–Free AcceleratorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Development and commissioning of a Monte Carlo photon beam model for Varian Clinac 2100EX linear acceleratorApplied Radiation and Isotopes, 2006
- The effect of flattening filter design on quality variations within an 8-MV primary x-ray beamMedical Physics, 1984